Matheson LeakHunter Plus 8066 Скачать руководство пользователя страница 3

1.0   Introduction 

The LeakHunter Plus, Model 8066, is a hand held instrument typically used for detecting 
leaks in gas plumbing joints, and for testing the integrity of sealed units using a tracer 
gas. The magnitude of the leak is indicated by a bargraph display and an audible tick. 

Operation can be one, or two handed. The audible tick and a peak hold mode allows 
operation in awkward situations where the display is not visible. The peak hold mode 
causes the display to retain the highest reading detected.  A nozzle is available for 
measuring large leaks by allowing the sample to be thirty times greater than with it off.  
An extended probe is supplied to facilitate access in difficult applications. 

An internal pump draws leakage gas and surrounding air in through the nozzle, then 
over the sensor that measures the thermal conductivity of the gas/air mixture. Thermal 
conductivity is defined as the rate of flow of heat through a medium. The gas/air mixture 
escapes from the instrument via four holes in its base. 

The LeakHunter Plus will indicate when a gas is present that has a thermal conductivity 
that differs from that of the ambient air on which it was zeroed.  The instrument is 
normally zeroed in ambient air comprising a mixture of approximately 78% nitrogen and 
21% oxygen. Hence, the instrument is not very sensitive to nitrogen or oxygen, but has 
a high sensitivity to helium and hydrogen since these have very different thermal 
conductivities from air. (See Section 8.)  A gas may have a higher or lower thermal 
conductivity than air, and thus produce a positive or negative reading on the display. 
The instrument has four gas group ranges to target the gases having thermal 
conductivities that differ from air by different degrees. 

The instrument can indicate a leak in terms of concentration (ppm), or flow rate (cfm, 
ml/sec or ml/min). Readings are approximateand relative; the instrument is intended to 
aid in the precise location of a leak, and to give an indication of its severity. Flow rate 
readings assume that all the leaking gas passes through the probe: If some of the 
leaking gas escapes to the surrounding air, without passing through the instrument, the 
flow reading will be inaccurate to that extent. 

The unit is battery operated using standard or rechargeable battery cells.  A power 
transformer is supplied for operating the unit from a 120VAC source. 

INT-0124-XX                     3

 

Содержание LeakHunter Plus 8066

Страница 1: ...LeakHunter Plus 8066 User Manual LeakHunter Plus LTR REVISION DATE BY A1 Release for Prod 11 02 95 CJC B1 Revise Spares List 05 23 96 BT INT 0124 XX...

Страница 2: ...anties shall not apply if the equipment has been repaired other than by Matheson or a service facility designated by Matheson or if this equipment has not been operated and maintained in accordance wi...

Страница 3: ...conductivity that differs from that of the ambient air on which it was zeroed The instrument is normally zeroed in ambient air comprising a mixture of approximately 78 nitrogen and 21 oxygen Hence th...

Страница 4: ...nge extension nozzle Probe clip and template AC power adapter and battery charger Battery User manual Cleaning wipe Extended probe The clip attaches permanently to the rear of the instrument The probe...

Страница 5: ...n in Figure 1 3 Peel off the paper that protects the adhesive on the back of the clip Assembly aid Probe clip Rear of case Figure 2 Probe clip template 4 Carefully position the clip in the position de...

Страница 6: ...ctable Functions Peak hold Zero Calibrate Gas type nozzle type Response time Approximately 1s Recovery time Approximately 1s Audio Signal proportional to leak rate Diagnostics Battery low indication D...

Страница 7: ...d in the same mode as it will be used 3 4 Peak hold Press PEAK HOLD to select and display the Peak Hold symbol on the screen In this mode the instrument shows the highest positive or negative value me...

Страница 8: ...zzle Present symbol The same operations should be carried out when attaching or removing the extended probe supplied as part of the 8066 3 8 Battery low This symbol becomes visible when the battery is...

Страница 9: ...and hold ZERO then press PEAK HOLD You will see the displayed exponent change with the change of sensitivity For instance with ml sec units successive exponents will be 5 most sensitive range 4 and 3...

Страница 10: ...are disabled 4 Present a standard leak or concentration of gas to the instrument 5 Set the display bargraph and exponent to read the standard leak flow rate or concentration use PEAK HOLD to increase...

Страница 11: ...pter is for indoor use only The battery charger is for indoor use with Ni Cad batteries only Do not attempt to recharge other types of battery as damage or injury may result Recharging time is 16 hour...

Страница 12: ...nd probe Instrument will not power up Depleted battery Replace battery or use AC power adapter Wrong connector used on AC power adapter Make sure smaller female connector is used on unit 7 2 Error ind...

Страница 13: ...05 6 0E 04 2 1E 08 1 6E 02 2 Neon 41 5 80E 05 3 5E 03 1 2E 07 9 3E 02 Xenon 41 5 81E 05 3 5E 03 1 2E 07 9 3E 02 3 R11 35 6 84E 05 4 1E 03 1 5E 07 1 0E 03 R12 30 7 90E 05 4 7E 03 1 7E 07 1 3E 03 R21 3...

Страница 14: ...Units Peak hold Battery low Leak rate bargraph mantissa Leak rate exponent Not used Default calibration in use Nozzle present LeakHunter Plus 8066 ZERO PEAK HOLD GAS GROUP 1 2 POWER INT 0124 XX 14...

Отзывы: