6.4.4 Adjusting the Lamp Alignment
After you install the lamp, position the lamp for maximum lamp energy. The flowchart in Figure
6-18 illustrates the procedure for positioning the lamp adjustment plate (Figure 6-16) for
maximum lamp energy.
Note: The detector does not have to be stabilized to perform this procedure.
Figure 6-18 Adjusting for Maximum Lamp Energy
To maximize lamp energy (see Figure 6-18):
1.
Turn on power for the CHDF 2000. The Power-On Diagnostics screen appears (see Figure
3-2).
Caution: To prevent the possibility of exposing your eyes to ultraviolet radiation, wear eye
protection that filters ultraviolet light and keep the lamp in the lamp housing during
operation.
2.
Make a note indicating the result (pass/fail) of the Calibrate UV power-on diagnostic.
3.
Choose Run diagnostics. The Run Diagnostics screen appears (see Figure 5-5).
4.
Choose Show Energy, AU,
λλλλ
. The current wavelength (
λ
) setting, the detected
absorbance (AU), and the current lamp sample (SMP) and reference energy (REF) values
appear (see Figure 5-6).
If the result of the Calibrate UV power-on diagnostic was Pass, go to step 10.
If the result of the Calibrate UV power-on diagnostic was Fail, continue with step 5.
Wait 5 seconds
After you replace
the lamp...
Turn power on
Did
power-on diagnostic
Calibrate UV
pass?
Choose diagnostic:
Show Energy, AU,
λ
Choose diagnostic:
Show Energy, AU,
λ
Use selector knob
to adjust
λ
to 656
Adjust height of lamp
adjustment plate to
maximize REF value
Rotate lamp
adjustment plate to
maximize REF value
Rotate lamp
adjustment plate to
maximize REF value
Adjust height of lamp
adjustment plate to
maximize REF value
Turn power off
Yes
No
Adjustment
complete