Next Screen - Selecting Next Screen brings up the third of the three configure system screens as
seen below in Figure 3.10
The third Configure System screen is displayed by selecting Next Screen in the second Configure
system screen. The third Configure System screen is shown below in Figure 3.10.
Figure 3.10 The third configure system screen.
When each of these commands is selected, the display changes so that a number may be entered.
The parameters are described and set as follows:
AU Zero Offset - Tells the instrument how many absorbance units of signal being measured by
the detector should be set equal to zero. Normally, this parameter should be left at zero. Once
this parameter is selected, the display changes to “mAU = :”. Rotate the selector knob to the
desired value, then push the knob to enter the value.
Eluant Temp - This parameter is used to calibrate the temperature probe on the CHDF 2000.
After selecting this parameter, rotate the selector knob until the current value of the eluant
temperature appears. Then push the knob to save the changes.
Auto Zero - Turns the autozeroing ability of the instrument on or off. When on, an autozero of
the detector baseline takes place each time an injection occurs. The autozero sets the detector
baseline to the mAU value defined in AU Zero Offset. Once selected, this parameter is turned on
or off by rotating the selector knob. Push the knob to record the change.
Serial Parameters - Sets the communication parameters that allow the CHDF 2000 to talk to a
personal computer. Once selected, display changes to “Baud Rate”. Rotate the selector knob the
appropriate value (normally 19200), then push the knob. This value must be the same as that set
in the software under system/configure system. Next, the display changes to “Echo” and is
followed by an “Off” or “On” (normally off). Rotate the selector knob to the desired choice, then
push the knob to save the settings.
Y Scale - Sets the maximum absorbance reading that can be displayed on the CHDF 2000 front
panel raw data graph during a fractionation. When selected, the display changes to “Y Axis AU