8
®
®
TROUBLESHOOTING (Continued)
If an error message is displayed, refer to the following list for possible corrective actions you can take.
If these do not correct the problem, contact your dealer.
ERROR MESSAGE
CAUSE
REMEDY
“Err 2”
Motor over-speed
1
Clear by pressing any key
2
Check for proper tube loading and pump operation
3
Return unit for repair if the error persists
“Err 3”
Overload
1
Clear by pressing any key
“Err 5”
2
Check for proper tube loading and pump operation
“Err 12”
3
Return unit for repair if the error persists
“Err 7”
Bad data Operator
1
Clear by pressing any key
parameters set to
2
Reprogram operator parameters
default values
3
Return unit for repair if the error persists
“Err 10”
Voltage out of range
1
Clear by pressing any key
“Err 11”
2
Check that AC line voltage is within specified
voltage ranges
3
Return unit for repair if AC line voltage is correct
and the error persists
“Err 13"
Over temperature
1
Check for heat sources or obstructions to cooling
“Err 14"
2
Check for proper tube loading and pump operation
3
Allow unit to cool Clear by pressing any key
4
Return unit for repair if no cause for overheating
is found and the error persists
All other errors
Internal error or failure
1
Clear (if possible) by turning power off and on
2
Return unit for repair if the error persists
CLEANING
Keep the drive enclosure clean with mild detergents. Do not immerse.
REPLACEMENT PARTS & ACCESSORIES
77500-24
Fuse—T6.3A 5 x 20 mm
07595-43
Footswitch
77300-32
Remote control cable, 25 ft (7.62 m)