5

®

®

REMOTE CONTROL (Model 77411-00 only)

Selectable input (4–20 mA, 0–10V DC )

±0.5% linearity control

±50V common mode range with respect to earth ground

Internal & External START/STOP; External CW/CCW via contact closure

REMOTE CONTROL SETUP

1.

Place the power switch in the off position.

2.

Connect the cable from the external remote control to the mating receptacle on the rear panel.

3.

Select operation from front panel potentiometer, external 4-20mA current source, or external 
0-10 V voltage source as follows:
a.

Press and hold the INT/EXT control for approximately 5 seconds until the present 
setting of the speed source is displayed. 

b.

Continuing to hold the INT/EXT control will cause the display to cycle through the three choices:
“4.20” indicating the 4-20 mA input, “0.10" indicating the 0-10 V input, or “Pot” indicating that the 
Speed Input Potentiometer will be used. When the key is released, the programmed choice will 
remain on the display for approximately 3 seconds as confirmation of the speed source choice 
made before returning to normal operation.

4.

To adjust the voltage or current scaling for other than zero to full scale:
a.

Press and hold the INT/EXT control (approximately 5 seconds) until the external speed 
source is displayed. (“4.20”, “0.10”, or “Pot”)

b.

Release the INT/EXT control and press the DIR control before the external speed 
source disappears from the display.

c.

The display will show “Lo” for about 3 seconds and then flash the current speed setting for the low set 
point (4 mA or 0 V). To change this setting, adjust the Speed Control Potentiometer to the percent of full
speed desired. To make no change, press the DIR  control a second time.

d.

After pressing the DIR control the second time, the display will show “Hi” and then flash 
the high (20 mA or 10 V) speed set point. This can be changed by adjusting the Speed 
Control Potentiometer, or left as is by pressing the DIR control again.

e.

The third press of the DIR control returns the drive to normal operation.

NOTE:

The 4-20 mA input, the 0-10 V input, and the 4-20 mA output are not scaled separately.

INTernal / EXTernal Operation

1.

The EXTernal mode of operation enables the PUMP READY output, the EXTERNAL START/STOP, EXTERNAL
CW/CCW, 4-20 mA, and 0-10 V inputs, while disabling the INTERNAL START/STOP input (e.g. footswitch) and
front-panel DIR and START controls. The pump speed is determined by the programmed choice of 4-20 mA, 
0-10 V, or front panel potentiometer inputs. The front panel STOP/START control overrides the EXTERNAL
START/STOP input to stop the drive.

2.

In the INTernal mode of operation, either the INTERNAL START/STOP input (e.g. footswitch) or the front panel
STOP/START control can start or stop the drive at the speed set by the front-panel SPEED POT.

Содержание I/P 77410-10

Страница 1: ...À DI CONTROLLO POMPA T1 T10 77411 00 pump drive Cole Parmer Instrument Co 1 800 MASTERFLEX 627 8373 U S and Canada only 11 847 549 7600 Outside U S 847 549 7600 Local www masterflex com Barnant Company 1 800 637 3739 U S and Canada only 11 847 381 7050 Outside U S 847 381 7050 Local www barnant com A 1299 1037 Edition 02 Model Nos N de modèle Modellnummer Número de modelo Modello nº 77410 10 77411...

Страница 2: ...priate measures to protect operator and equipment Turn drive off before removing or installing tubing Fingers or loose clothing could get caught in drive mechanism To avoid electrical shock the power cord protective grounding conductor must be connected to ground Not for operation in wet locations as defined by EN 61010 1 CAUTIONS Do not stack drives Keep 3 minimum distance around and above drive ...

Страница 3: ...p heads One Masterflex High Performance I P pump head may be driven CONTROL DISPLAY FUNCTIONS Press keys to activate function A B C A D B C A STOP START Start Stop pump B DIRECTION To change motor direction C SPEED Selects output speed from 5 to 100 Full Scale 33 to 650 RPM Clockwise rotation increases speed D INT EXT Model 77411 00 only INT for internal control EXT for remote current or voltage c...

Страница 4: ...DIR on power up After five 5 seconds display will read ON factory default or OFF Hold the DIR key for an additional 3 seconds and the display will toggle to the other option every 3 seconds Release the DIR key when the desired mode is being displayed After releasing the key ON or OFF will remain displayed for about 3 seconds and then the normal power on sequence will begin When ON is selected driv...

Страница 5: ...ase the INT EXT control and press the DIR control before the external speed source disappears from the display c The display will show Lo for about 3 seconds and then flash the current speed setting for the low set point 4 mA or 0 V To change this setting adjust the Speed Control Potentiometer to the percent of full speed desired To make no change press the DIR control a second time d After pressi...

Страница 6: ...3 BLACK BROWN VIOLET N C N C PINK RED GREEN RED BLACK PUMP READY N O CONTACT EXTERNAL START STOP EXTERNAL CW CCW INTERNAL START STOP RED RESERVED N C TAN RESERVED OUTPUT 4 20mA INPUT 0 10V INPUT 4 20mA NOTE Colors are those of Remote Cable Catalog number 77300 32 PUMP READY N C CONTACT Figure 2 ...

Страница 7: ...ne 3 Check connection of power cord 4 Check the line cord for continuity and replace if defective 5 Return for servicing B Motor does not rotate B1 Defective remote 1 Place power switch in off position Display lights control 2 Check that remote cable connector is inserted fully into the receptacle Fig 2 and 3 3 If motor still does not rotate select INTernal with the INT EXT control and press the S...

Страница 8: ... unit for repair if the error persists Err 10 Voltage out of range 1 Clear by pressing any key Err 11 2 Check that AC line voltage is within specified voltage ranges 3 Return unit for repair if AC line voltage is correct and the error persists Err 13 Over temperature 1 Check for heat sources or obstructions to cooling Err 14 2 Check for proper tube loading and pump operation 3 Allow unit to cool C...

Страница 9: ... Vrms or 2 6A 230 Vrms Remote Inputs Internal External START STOP Contact closures Model 77411 00 only External CW CCW Contact closure Voltage input 0 10V DC 10 kΩ Current input 4 20 mA 250 Ω CONSTRUCTION Dimensions L x W x H 14 0 x 10 x 9 1 35 cm x 25 cm x 23 cm Weight 22 lb 10 kg Enclosure rating IP55 per IEC 60529 ENVIRONMENT Temperature operating 0 to 40 C 32 to 104 F Temperature storage 25 to...

Страница 10: ... 8373 U S and Canada only 11 847 549 7600 Outside U S 847 549 4600 Local FAX 847 247 2929 U S and Canada only 11 847 549 1700 Fax outside U S www masterflex com e mail techinfo coleparmer com 28W092 Commercial Ave Barrington Illinois U S A 60010 2392 1 800 637 3739 U S and Canada only 11 847 381 7050 Outside U S 847 381 7050 Local FAX 11 847 381 7053 Fax outside U S 847 381 7053 Local Fax www barn...

Отзывы: