2.  Place the PURO PRO in desired position, immediately following 

any prescribed pretreatment, and far enough from walls and other 
obstructions to allow for servicing the unit.  Note that 14 inches of 
horizontal clearance are required on the open end for changing 
the filters. 

3.  Place the PURO PRO within reasonable access to a grounded 

115V/60 HZ  circuit and a legal drain line connection.  

4.  This system will weigh about 70 pounds when full of water.   
 

NOTE:

 When changing membranes, water will spill on the floor, so 

installing near a floor drain or unfinished floor area is preferred. 

 
PURO PRO Hydraulic Connections (Figure 5): 

 

1.  Pipe Puro Pro into the service lines, the inlet is ½ ” NPT and the 

outlet is ½” OD tube compression.  We recommend using an 
inlet, outlet, and bypass valve in installation piping for future 
servicing of unit.  Always follow local plumbing codes when 
installing our water treatment equipment.  When using PVC 
make sure to follow the proper primer and solvent instructions. 

2.  Connect a pressurized feed water source to the inlet solenoid.  

The solenoid has a ½ -inch NPT connection.  This water source 
should be consistently pressurized, either from a well pump 
pressure (bladder) tank or from a municipal water source. 

3.  Install ½” OD tube from the Puro Pro check valve to the inlet of 

your atmospheric tank. 

4.  Install a ½ ” drain line from the ½ ” NPT drain connection to a 

legal air gap.  Follow local plumbing codes. 

 

 

Содержание PURO PRO 1800 SC Series

Страница 1: ...OPERATION MANUAL PURO PRO 1800 SC Series Modular Reverse Osmosis System MASTER Water Conditioning Corp www masterwater com September 2012 Version...

Страница 2: ...nections 9 10 PURO PRO Power Switch Operation 10 11 PURO PRO Float Switch Operation 11 12 Filling the PURO PRO with Water and Start Up 13 PURO PRO Operation and Maintenance 13 14 Production Data Recor...

Страница 3: ...d Solids TDS from water The system consists in six 6 high performance reverse osmosis membranes The system is designed to operate with a pressurized 40 to 90 psi feed and a vented storage tank for col...

Страница 4: ...be correct Avoid using outlets that are switch controlled 2 Maximum amperage required is 5 amps 3 Make sure the electrical service provides power 24 hours per day We recommend installing a surge prot...

Страница 5: ...3 Figure 1 Typical PURO PRO Installation...

Страница 6: ...eservative 5 Ensure that the O Ring s on each housing and membrane has Silicone only Lubricant Do not use petroleum lubricants 6 Install two of the membranes in the top row 7 To install the membranes...

Страница 7: ...rane Placing Membrane In Manifold Inserting Membrane Housing Into Manifold 8 Repeat with the second membrane in the middle row 9 Install the remaining membranes using the procedure from steps 7 and 8...

Страница 8: ...6 Figure 2 Manifold Figure 3 Housings Pre Filter Manifold Post Filter Manifold 4 Membrane Manifolds Housing O Ring...

Страница 9: ...O Positioning 1 Wall mount the PURO PRO There are holes drilled into the plastic housing to be used for wall mounting If using the PURO PRO on the ground make sure to place a support under the bottom...

Страница 10: ...nnections Figure 5 1 Pipe Puro Pro into the service lines the inlet is NPT and the outlet is OD tube compression We recommend using an inlet outlet and bypass valve in installation piping for future s...

Страница 11: ...when in the horizontal floating position See Figure 1 2 Connect the float switch to the power box with the electrical connector provided See Figure 6 3 The inlet solenoid is already connected to the...

Страница 12: ...float switch is in the vertical position Whenever the storage tank is full the float switch is in the horizontal floating position the solenoid will be closed PURO PRO Power Switch Operation 1 The pow...

Страница 13: ...is closed power is passed and the solenoid valve is opened allowing the Modular Pro to produce treated water 2 This function can be tested by holding the float adjusting the orientation and listening...

Страница 14: ...12 Low Water Switch On High Water Switch Off WEIGHT METHOD Low Water Switch On High Water Switch Off ANCHOR POINT METHOD WEIGHT DEVICE...

Страница 15: ...sure that the water level is as required Adjust the float rotation point as necessary PURO PRO Operation and Maintenance 1 To ensure proper maintenance it is recommended that a production data sheet b...

Страница 16: ...14 6 If the module frame structure becomes loose tighten the FRP Bolts located at the frame corners...

Страница 17: ...duct Quality Action Taken Product measurement criteria is selected based upon the configuration of the system For example an appropriate criteria may be gallons per hour or inches of tank height per h...

Страница 18: ...have become severely fouled Replace and re start Symptom PURO PRO has greatly reduced production Possible Cause Solution Water pressure is decreased below 60 psi Restore water pressure Drain control...

Страница 19: ...oid has failed Replace inlet solenoid Symptom Storage Tank Overflows Possible Cause Solution Float Switch has failed Replace float switch Inlet Solenoid has failed Replace the inlet solenoid PURO PRO...

Страница 20: ...freezing or hot water All replacement filters and membranes must be Master Water Conditioning filters and membranes or the warranty is void If the unit is altered by anyone other than Master Water Co...