Facility
Requirements
17
Need help? Contact [email protected]
Site Preparation
Manual
North America connection diagram (
For all country with 3ph 208/220 VAC)
Figure 8-North America connection diagram
Uninterruptible Power Supply (UPS) Requirements
The customer is responsible for providing a UPS system to support the PC, monitor and
controllers, with the following specifications:
For Old computer
• Voltage output: 230 VAC
• 24VDC_Power-Supply - 240W
• Monitor - 30W
• PC – 24VDC-60W
For New computer
• Voltage output: 230 VAC
• 24VDC_Power-Supply - 240W
• Monitor - 30W
• PC – 230VAC-60W
R
S
T
R1
R2
S1
S2
T1
T2
N
G
Air connection
European type
R1 S1T1 G
Dongan three phase
Transformer
Cat NO: 63-45-815SH
Primary 208VAC
Secondary 380Y/220
Δ