Chapter 8: Troubleshooting
www.masimo.com
46
Masimo
Message
Action
Gel Bridging Detected
1.
Clean any gel that has leaked outside of the electrodes on
the patient's forehead.
2.
Confirm that all electrodes of the sensor are properly
connected.
3.
The sensor may need to be replaced.
Low SIQ message
displayed (Low signal
quality)
Re-apply sensor. See
Directions for Use
for Sensor.
Adjusting Electrodes
The following scenarios may indicate an issue with electrodes or cabling and may be resolved
by improving patient-electrode contact:
•
If Root displays an electrode status icon in any color other than green.
•
If a signal is noisy or high in amplitude and appears different from the other
channels.
To improve electrode contact
Note:
After making any electrode adjustments, wait two (2) to three (3) seconds for SedLine
to update. It is important to check the electrodes in the order listed below:
Icon Color
Action
Green
No electrode adjustment necessary.
Yellow
Minor electrode adjustment may be required.
Red
Gently push/wiggle affected electrodes until all are yellow and/or green.