SedLine Sedation Monitor
Chapter 5: Menu Options
www.masimo.com
36
Masimo
Options
Description
Alarm
Priority
Factory
Default
Selectable
Settings
Silence
Duration
Length of time that the
audible alarm remains
silenced
NA
2 min
30 sec, 1, 2, or 5
min
Audible
Alarms*
Disables audible alarm
NA
Off
On or Off
* SedLine has the ability to control alarm priority based on the clinical use case. When
Audible Alarms are set to Off, the alarm priority is Low. When Audible Alarms are set to On,
the alarm priority is Medium.
Trends
Options
Description
Factory
Default Selectable Settings
Y-Axis Max Highest PSi value that can be displayed 100
5 to 100 in steps of 5
Y-Axis Min
Lowest PSi value that can be displayed 0
0 to 95 in steps of 5
Threshold
Max
Upper limit of target PSi range (green
band on PSi Trend display)
50
5 to 100 in steps of 5
Threshold
Min
Lower limit of target PSi range (green
band on PSi Trend display)
25
0 to 95 in steps of 5
DSA
About DSA
An informational read-only screen with definition of DSA:
The Density Spectral Array (DSA) is a color representation of the power of the EEG between 0
and 30/40 Hz showing activity in the right and left areas of the brain using spectral edge
frequency as the indices. The DSA represents the last 20 minutes of R and L frontal-prefrontal
EEG activity tracked by the 95% Spectral Edge Frequency.
Содержание sedline
Страница 1: ...Operator s Manual SedLine Sedation Monitor...
Страница 2: ......
Страница 4: ......
Страница 8: ......
Страница 10: ......
Страница 14: ......
Страница 16: ......
Страница 20: ......
Страница 24: ......
Страница 46: ......
Страница 62: ......
Страница 65: ......
Страница 66: ...www masimo com 35548 LAB 7373G 0619...