SedLine Sedation Monitor
Chapter 5: Menu Options
www.masimo.com
38
Masimo
Trends
Options
Description
Factory
Default
Selectable Settings
Y-Axis
Max
Highest EMG value that can be
displayed
100%
5% to 100% in 5%
increments
Y-Axis
Min
Lowest EMG value that can be
displayed
0%
0% to 95% in 5 %
increments
Additional Settings*
Options
Description
Factory
Default
Selectable
Settings
EMG
Sensitivity
Sets the sensitivity level for EMG
detection.
Standard
Standard or
Maximum
* Available on SedLine with software V232x or higher.
SR
About
An informational read-only screen with definition of SR:
Suppression Ratio (SR) is a measure of how much the electrical activity of the frontal and
pre-frontal cortex of the brain is suppressed.
Alarms
Options
Description
Alarm
Priority
Factory
Default
Selectable
Settings
High Limit
Upper limit that triggers an
alarm
Low/Medium 10
2 to 99 in steps
of 1, or Off
High Caution
Range
Value below High Limit to
trigger caution light
NA
Off
Off, or 1 to 10 in
steps of 1
Silence
Duration
Length of time that the
audible alarm remains
silenced
NA
2 min
30 sec, 1, 2, or 5
min
Trends
Содержание sedline
Страница 1: ...Operator s Manual SedLine Sedation Monitor...
Страница 2: ......
Страница 4: ......
Страница 8: ......
Страница 10: ......
Страница 14: ......
Страница 16: ......
Страница 20: ......
Страница 24: ......
Страница 46: ......
Страница 62: ......
Страница 65: ......
Страница 66: ...www masimo com 35548 LAB 7373G 0619...