Rad-67
Chapter 6: Alerts and Messages
www.masimo.com
65
Masimo
Message
Potential Causes
Next Steps
(Pulse CO-Ox)
Interference
Detected
High intensity light
(pulsating strobe lights,
excessive ambient light
sources such as surgical
lights or direct sunlight)
or other monitor displays.
Incorrect monitor line
frequency setting (Hz).
Place a Masimo Optical Light
Shield over the sensor.
Adjust the Line Frequency to the
correct Hz setting. See
Device
Settings
on page 47.
Low Perfusion
Signal strength is too
weak.
Select different sensor site that is
better perfused. See
Troubleshooting Measurements
on page 67.
Low Signal
Stability
Motion during SpHb
measurement.
Keep measurement site stable.
See
Troubleshooting
Measurements
on page 67.
Demo Mode
Device is in demo mode
setting.
Disable demo mode prior to
patient spot-check
measurements.
SpO
2
Only
Mode
Occurs during an
unsuccessful SpHb sensor
initialization/pulse search
routine or during
measurement.
See the Directions for Use
provided with your sensor.
Use a Masimo light shield to
cover the sensor and adjust the
sensor.
Ensure proper sensor application.
Replace sensor.
Low SpHb SIQ
Indicates low signal
quality of SpHb
measurement.
The SpHb reading is
outside of the accuracy
range.
Ensure proper sensor application.
Replace the sensor.
See
Chapter 8: Specifications
on
page 75.
Содержание Rad-67 Pulse CO-Oximeter
Страница 1: ...Operator s Manual Rad 67 Pulse CO Oximeter Spot Check...
Страница 2: ......
Страница 4: ......
Страница 8: ......
Страница 10: ......
Страница 12: ......
Страница 20: ......
Страница 58: ......
Страница 76: ......
Страница 94: ......
Страница 100: ......
Страница 104: ......
Страница 105: ......
Страница 106: ...www masimo com 300362 LAB 10262A 0419...