Rad-67
Chapter 5: Spot-Check
www.masimo.com
62
Masimo
Limits of Agreement (LoA) Gauge
Rad-67 provides an LoA gauge with white bars to indicate the upper and lower 95% limits of
agreement for the displayed SpHb measurement. In the gauge, the dot indicates where the
SpHb measurement falls across the display range. LoA is an interval that provides an
estimate of accuracy. 95% of the time, the difference between the SpHb value and a gold-
standard reference analyzer lie within this range.
In the example shown, SpHb displayed is
within the accuracy range and the LoA gauge is
displayed.
This example shows when the SpHb
measurement is outside of the accuracy range
(see
SpHb Accuracy Data
on page 77).
The
SpHb out of validated range
message
displays. The LoA gauge is disabled.
See
Troubleshooting Measurements
on page
If low SIQ is encountered while attempting to
read SpHb, SpHb is displayed as dashes "--",
the LoA gauge is disabled and
low SpHb SIQ
displays.
See
Troubleshooting Measurements
on page