Rad-67
Chapter 4: Operation
www.masimo.com
36
Masimo
Connect the Rad-67 to AC power to prevent the device from powering off and to charge the
battery. When connected to power, the AC Power Indicator icon will be displayed.
Touch the Battery Charge Status Indicator icon to view battery details. See
Rad-67 Battery
on page 52.
Low Power Mode
To conserve the Rad-67 battery, the device automatically dims the display after
approximately 10 seconds, and turns the display off after another 10 seconds, when the
following criteria are met:
The device is not currently performing a spot-check measurement on a patient.
AND
There is no user interaction with the device display.
Touch the display to return it to its full brightness.
About the Action Menu
To expand the
Action Menu
, select the arrow in the upper right corner of the window.
The Action Menu allows quick access to the following settings directly from the
Main Screen
:
Sensitivity
- Selecting this option cycles through the available sensitivity modes:
APOD, NORM and MAX. See
Sensitivity Modes Overview
on page 36.
Note:
After approximately 10 seconds without interaction, the
Action Menu
will retract.
Sensitivity Modes Overview
Three sensitivity levels enable a clinician to tailor the response of Rad-67 to the needs of
the particular patient situation. Sensitivity Modes are accessed through the
Action Menu
.
See
About the Action Menu
on page 36.
The sensitivity levels are as follows:
NORM (Normal Sensitivity)
NORM is the recommended sensitivity mode for patients who are experiencing
some compromise in blood flow or perfusion. It is advisable for care areas where
patients are observed frequently, such as an intensive care unit (ICU).
APOD® (Adaptive Probe Off Detection® Sensitivity)
APOD is the recommended sensitivity mode for situations which there is a high
probability of the sensor becoming detached. It is also the suggested mode for
care areas where patients are not visually monitored continuously. This mode
Содержание Rad-67 Pulse CO-Oximeter
Страница 1: ...Operator s Manual Rad 67 Pulse CO Oximeter Spot Check...
Страница 2: ......
Страница 4: ......
Страница 8: ......
Страница 10: ......
Страница 12: ......
Страница 20: ......
Страница 58: ......
Страница 76: ......
Страница 94: ......
Страница 100: ......
Страница 104: ......
Страница 105: ......
Страница 106: ...www masimo com 300362 LAB 10262A 0419...