Chapter 4: Operation
www.masimo.com
41
Masimo
To access any of the available parameter setting screens:
1.
From the
Parameter Settings
screen, to access the desired parameter, swipe the
on-screen icons left or right.
2.
Touch the icon of the desired parameter.
See
SpO2 Settings
on page 41.
See
PR Settings
on page 42.
See
SpHb Settings
on page 42.
See
Pi Settings
on page 42.
SpO2 Settings
Allows access to any of the following options:
Additional Settings for SpO2
on page 41
About Parameter Information
on page 43
Additional Settings for SpO2
From the
Additional Settings
screen, change any of the following options:
Options
Description
Factory
Default
Settings
User Configurable
Settings
Averaging
Time*
The length of time over which the
system calculates the average of all
data points.
8 seconds
2-4, 4-6, 8, 10, 12, 14,
or 16 seconds**
FastSat
See
FastSat Overview
on page 41.
Off
Off or On
* With FastSat the averaging time is dependent on the input signal.
** For the 2 and 4 second settings the averaging time may range from 2-4 and 4-6 seconds,
respectively.
FastSat Overview
FastSat enables rapid tracking of arterial oxygen saturation changes. Arterial oxygen
saturation data is averaged using pulse oximeter averaging algorithms to smooth the trend.
When Rad-67 is set to FastSat
On
, the averaging algorithm evaluates all saturation values,
providing an averaged saturation value that is a better representation of the patient’s
current oxygenation status. With FastSat set to On, the averaging time is dependent on the
input signal.
Содержание Rad-67 Pulse CO-Oximeter
Страница 1: ...Operator s Manual Rad 67 Pulse CO Oximeter Spot Check...
Страница 2: ......
Страница 4: ......
Страница 8: ......
Страница 10: ......
Страница 12: ......
Страница 20: ......
Страница 58: ......
Страница 76: ......
Страница 94: ......
Страница 100: ......
Страница 104: ......
Страница 105: ......
Страница 106: ...www masimo com 300362 LAB 10262A 0419...