Chapter 8: Service and Maintenance
The following chapter contains information about cleaning, battery operation, performance
verification, service, repair, and warranty.
Cleaning
To clean the device:
1.
Disconnect the AC Power Supply and ensure the sensor is not applied to the
patient.
2.
Turn off the device and remove the batteries. See
Battery Installation and
Replacement
on page 57.
3.
Wipe the outer surfaces using a dampened soft cloth with a mild detergent and
warm water solution or one of the recommended cleaning solutions twice or until
the surfaces are free of any visible residue.
CAUTION:
Do not allow liquids to enter the interior of the device.
4.
Dry the device thoroughly prior to using on a patient.
The surfaces of the Rad-5/5v may be cleaned with the following solvents or cleaning agents:
70% Isopropyl Alcohol (IPA)
10% bleach/water solution (~0.5% sodium hypochlorite solution)
Up to 55% alcohol/0.5% quaternary ammonium chloride solution, such as Ecolab
Asepti-Wipe
®
II Germicidal Wipes or PDI Super Sani-Cloth
®
Germicidal Wipes
Accelerated Hydrogen Peroxide
®
Oxivir
®
Tb
Cidex
®
Plus (3.4% glutaraldehyde)
CAUTION:
Do not use undiluted bleach (5% - 5.25% sodium hypochlorite) or any cleaning
solution other than those recommended here because permanent damage to the device may
occur.
CAUTION:
To prevent damage, do not soak or immerse the device in any liquid solution.
CAUTION:
Do not sterilize by irradiation, steam, and autoclave or ethylene oxide.
Battery Installation and Replacement
WARNING:
Use only alkaline batteries. Use of non alkaline batteries may affect the accuracy
of the battery level meter.
WARNING:
Use of batteries with a cell voltage of more than 1.5V could cause damage to the
Device.
WARNING:
Remove batteries if unit is not to be used for some time.
1.
Remove the battery cover by depressing the small rectangular button at the
bottom of the cover. See
Back View
on page 21.
2.
Slide the cover down off the bottom of the device.
Содержание Rad-5
Страница 1: ...Operator s Manual Rad 5 5v Pulse Oximeter...
Страница 2: ......
Страница 4: ...www masimo com 2 Masimo 2020 Masimo Corporation...
Страница 8: ......
Страница 16: ......
Страница 20: ......
Страница 40: ......
Страница 46: ......
Страница 48: ......
Страница 68: ......
Страница 69: ......
Страница 70: ...www masimo com 33606 LAB 5724H 0920...