Rad-5/5v
Chapter 7: Specifications
www.masimo.com
51
Masimo
Equipment Classification per IEC 60601-1
Degree of Protection of
Electrical Shock
Type BF-Applied Part
Protection against
harm from liquid
ingress
IP22, Protection from ingress of particulates > than 12.5 mm and
protection against vertically falling water drops when enclosure
is tilted at 15 degrees.
Mode of Operation
Rad-5
Continuous Operation
Rad-5v
Not Continuous (Spot Check)
Guidance and Manufacturer's Declaration - Electromagnetic
Emissions
Guidance and Manufacturer's Declarations - Electromagnetic Emissions
The ME Equipment is intended for use in the electromagnetic environment specified
below. The customer or the user of the ME Equipment should assure that it is used in such
an environment.
Emission Test Compliance Electromagnetic Environment - Guidance
RF Emissions
CISPR 11
Group 1
ME Equipment uses RF energy only for its internal function.
Therefore, its RF emissions are very low and are not likely to
cause any interference in nearby electronic equipment.
RF Emissions
CISPR 11
Class B
Suitable for use in all establishments, including domestic
environments.
Harmonic
emissions
IEC 61000-3-
2
N/A
Voltage
fluctuations/
flicker
emissions
N/A
Содержание Rad-5
Страница 1: ...Operator s Manual Rad 5 5v Pulse Oximeter...
Страница 2: ......
Страница 4: ...www masimo com 2 Masimo 2020 Masimo Corporation...
Страница 8: ......
Страница 16: ......
Страница 20: ......
Страница 40: ......
Страница 46: ......
Страница 48: ......
Страница 68: ......
Страница 69: ......
Страница 70: ...www masimo com 33606 LAB 5724H 0920...