Chapter 4: Operation
www.masimo.com
32
Masimo
Note:
Turning trend off will not erase trend memory. You may turn trending off and still
retrieve the trend data using TrendCom.
Turning the Rad-5 off or replacing the batteries will not erase the trend data.
Turn trending off before storing the device for any length of time.
Trend Data Format
After a successful download of the trend data, an ".out" file is created containing the trend-
dump information in ASCII delimited format. The format is defined in the following table.
Parameter
Specification
Date
MM\DD\YY
Time
HH:MM:SS
SpO
2
001 to 100, or "---" meaning parameter not available
Pulse rate
001 to 240, or "---" meaning parameter not available
Perfusion
Index
00.00 to 20.00
Exception
Messages
The exceptions are displayed as a 3 digit, ASCII encoded, hexadecimal
value. The binary bits of the hexadecimal value are encoded as follows:
000 = Normal operation; no exceptions
001 = No Sensor
002 = Defective Sensor
004 = Low Perfusion
008 = Pulse Search
010 = Interference
020 = Sensor Off
040 = Ambient Light
080 = Unrecognized Sensor
100 = reserved
200 = reserved
400 = Low Signal IQ
800 = Masimo SET*
* This flag means the algorithm is running is full SET mode. It requires a SET sensor and
needs to acquire some clean data for this flag to be set.
Содержание Rad-5
Страница 1: ...Operator s Manual Rad 5 5v Pulse Oximeter...
Страница 2: ......
Страница 4: ...www masimo com 2 Masimo 2020 Masimo Corporation...
Страница 8: ......
Страница 16: ......
Страница 20: ......
Страница 40: ......
Страница 46: ......
Страница 48: ......
Страница 68: ......
Страница 69: ......
Страница 70: ...www masimo com 33606 LAB 5724H 0920...