Rad-5/5v
Safety Information, Warnings and Cautions
www.masimo.com
12
Masimo
CAUTION:
When patients are undergoing photodynamic therapy they may be sensitive to
light sources. Pulse oximetry may be used only under careful clinical supervision for short
time periods to minimize interference with photodynamic therapy.
CAUTION:
High ambient light sources such as surgical lights (especially those with a xenon
light source), bilirubin lamps, fluorescent lights, infrared heating lamps, and direct sunlight
can interfere with the performance of the sensor.
CAUTION:
To prevent interference from ambient light, ensure that the sensor is properly
applied, and cover the sensor site with opaque material, if required. Failure to take this
precaution in high ambient light conditions may result in inaccurate measurements.
CAUTION:
If the Low Perfusion message is frequently displayed, find a better perfused
monitoring site. In the interim, assess the patient and, if indicated, verify oxygenation
status through other means.
CAUTION:
To minimize radio interference, other electrical equipment that emits radio
frequency transmissions should not be in close proximity to Rad-5/5v.
Note:
Cables and sensors are provided with X-Cal™ technology to minimize the risk of
inaccurate readings and unanticipated loss of patient monitoring. Refer to the Cable or
Sensor Directions for Use for the specified duration of patient monitoring time.
Note:
Physiological conditions that result in loss of pulsatile signal may result in no SpO
2
readings.
Note:
A functional tester cannot be used to assess the accuracy of Rad-5/5v.
Note:
Additional information specific to the Masimo sensors compatible with Rad-5/5v,
including information about parameter/measurement performance during motion and low
perfusion, may be found in the sensor's directions for use (DFU).
Cleaning and Service Warnings and Cautions
WARNING:
Do not attempt to remanufacture, recondition or recycle the Rad-5/5v as these
processes may damage the electrical components, potentially leading to patient harm.
CAUTION:
Only perform maintenance procedures specifically described in the manual.
Otherwise, return the Rad-5/5v for servicing.
CAUTION:
Do not touch, press, or rub the display panels with abrasive cleaning compounds,
instruments, brushes, rough-surface materials, or bring them into contact with anything
that could scratch the display.
CAUTION:
Do not use petroleum-based or acetone solutions, undiluted bleach (5% - 5.25%
sodium hypochlorite solution), or other harsh solvents. These substances affect the device’s
materials and instrument failure can result.
CAUTION:
Do not use different cleaning solutions to clean the same instrument.
Unpredictable chemical reactions that may damage the instrument may occur.
CAUTION:
Do not immerse the instrument in any liquid solution.
CAUTION:
Do not sterilize by irradiation, steam autoclave, or ethylene oxide.
Содержание Rad-5
Страница 1: ...Operator s Manual Rad 5 5v Pulse Oximeter...
Страница 2: ......
Страница 4: ...www masimo com 2 Masimo 2020 Masimo Corporation...
Страница 8: ......
Страница 16: ......
Страница 20: ......
Страница 40: ......
Страница 46: ......
Страница 48: ......
Страница 68: ......
Страница 69: ......
Страница 70: ...www masimo com 33606 LAB 5724H 0920...