NomoLine ISA CO2
Chapter 5: Specifications
www.masimo.com
26
Masimo
Gas or Vapor
Gas Level
CO
2
CO (Carbon Monoxide)
5
1 vol%
_
1
NO (Nitrogen Monoxide)
5
0.02 vol%
_
1
O
2
5
100 vol%
_
2
Note 1:
Negligible interference, effect included in the specification “Accuracy, all
conditions” above.
Note 2:
Negligible interference with N
2
O / O
2
concentrations correctly set, effect included
in the specification “Accuracy, all conditions” above.
Note 3:
Interference at indicated gas level. For example, 50 vol% Helium typically
decreases the CO
2
readings by 6%. This means that if measuring on a mixture containing 5.0
vol% CO
2
and 50 vol% Helium, the actual measured CO
2
concentration will typically be
(1-0.06) * 5.0 vol% = 4.7 vol% CO
2
.
Note 4:
According to the EN ISO 80601-2-55 standard.
Note 5:
In addition to the EN ISO 80601-2-55 standard.
Symbols
The following symbols may appear on the product or product labeling:
Symbol
Description
Symbol
Description
Follow instructions for use
Consult instructions for use
Separate collection for electrical
and electronic equipment
(WEEE)
Mark of conformity to European
medical device directive
93/42/EEC
Protection against splashing of
water from all directions
ETL Intertek certification
See
Declarations on Page 1
for
certifications
Caution:
Federal (USA) law
restricts this device to sale by or
on the order of a physician
Defibrillation-proof. Type BF
applied part
Biohazardous Waste
Serial number
Содержание NomoLine ISA CO2
Страница 1: ...Operator s Manual NomoLine ISA CO2 Sidestream Gas Analyzer...
Страница 2: ......
Страница 4: ......
Страница 8: ......
Страница 10: ......
Страница 14: ......
Страница 18: ......
Страница 20: ......
Страница 30: ......
Страница 36: ......
Страница 39: ......
Страница 40: ...www masimo com 300106 LAB 10152A 0219...