Chapter 4: Troubleshooting
Troubleshooting NomoLine ISA CO2
Symptom
Possible Cause
Correction
NomoLine ISA
CO2 values are
not displayed on
host monitor
•
NomoLine ISA CO2
requires a power cycle
•
Host device is not
configured correctly or is
incompatible with
NomoLine ISA CO2
•
Disconnect and reconnect the
NomoLine ISA CO2 connector at
the host device.
•
Verify that the host device is
correctly configured and
compatible with NomoLine ISA
CO2.
LEGI indicator is
dark (indicator is
off)
•
NomoLine ISA CO2 is not
plugged in to the host
monitor connector
•
Sampling line not plugged
in to NomoLine ISA CO2
completely
•
Internal error
•
Verify NomoLine ISA CO2
connection to the host monitor.
•
Disconnect and reconnect the
sampling line to NomoLine ISA
CO2.
•
The NomoLine ISA CO2 requires
service. See Service and Return
Procedure.
LEGI indicator
blinking red
•
Sampling line obstructed
(occlusion)
•
Inspect sampling line for
occlusion (blockage, kink in line).
•
Replace sampling line. See
Sampling Line Replacement
on
page 14.
LEGI indicator
steady red
•
NomoLine ISA CO2
requires a power cycle
•
Internal error
•
Disconnect and reconnect the
NomoLine ISA CO2 connector at
the host device.
•
If persistent, the NomoLine ISA
CO2 requires service. See Service
and Return Procedure.
•
A hardware error is
generated during the
warm up period after being
stored at −40°C
Содержание NomoLine ISA CO2
Страница 1: ...Operator s Manual NomoLine ISA CO2 Sidestream Gas Analyzer...
Страница 2: ......
Страница 4: ......
Страница 8: ......
Страница 10: ......
Страница 14: ......
Страница 18: ......
Страница 20: ......
Страница 30: ......
Страница 36: ......
Страница 39: ......
Страница 40: ...www masimo com 300106 LAB 10152A 0219...