Chapter 4: Troubleshooting
www.masimo.com
20
Masimo
Symptom
Possible Cause
Correction
Gas readings are
questionable
•
Incorrect sampling line
used
•
Sampling line attached
incorrectly
•
Sampling line leaking
•
Internal error
•
Use the correct sampling line with
NomoLine ISA CO2.
•
Disconnect and reconnect the
sampling line to NomoLine ISA
CO2.
•
Follow the Directions for Use to
properly connect the sampling
line.
•
Check sampling line for leaks.
•
Replace sampling line. See
Sampling Line Replacement
on
page 14.
•
If persistent, the NomoLine ISA
CO2 requires service. See Service
and Return Procedure.