Safety Information, Warnings and 

Cautions 

CAUTION

: NomoLine ISA CO2 is to be operated by, or under the supervision of, qualified 

personnel only. Read this Operator's Manual, accessories directions for use, all precautionary 

information, and specifications before use.

 

Refer to the host device operator's manual or 

user's guide for additional safety information, warnings and cautions.   

 

Safety Warnings and Cautions 

WARNING:

 Do not use NomoLine ISA CO2 if it appears or is suspected to be damaged. 

Damage to the device can result in exposed electrical circuits that may cause patient harm. 

WARNING:

 Do not adjust, repair, open, disassemble, or modify the NomoLine ISA CO2. 

Damage to the device may result in degraded performance and/or patient injury. 

WARNING:

 Do not start or operate the NomoLine ISA CO2 unless the setup was verified to be 

correct. Improper set-up of this device may result in degraded performance and/or patient 

injury. 

WARNING:

 Do not place the NomoLine ISA CO2 or accessories in any position that might 

cause it to fall on the patient. 

WARNING:

 Only use Masimo authorized devices with NomoLine ISA CO2. Using 

unauthorized devices with NomoLine ISA CO2 may result in damage to the device and/or 

patient injury. 

WARNING:

 Carefully route the sampling line to reduce the risk of patient entanglement or 

strangulation. 

WARNING:

 Do not lift the NomoLine ISA CO2 by the NomoLine capnography sampling line as 

it could disconnect from the NomoLine ISA CO2, causing the device to fall on the patient. 

WARNING:

 Do not use the NomoLine ISA CO2 during magnetic resonance imaging (MRI) or 

in an MRI environment. 

WARNING:

 Only use sample lines intended for anesthetic agents if N

2

O and/or anesthetic 

agents are being used. 

WARNING:

 Do not re-use disposable single-patient use NomoLine Family sampling lines due 

to the risk of cross contamination. 

WARNING:

 Do not use the NomoLine Infant/Neonate Airway Adapter Sets for 

adults/pediatrics as they may cause excessive flow resistance (0,7 ml dead space). 

WARNING:

 Do not use the NomoLine Adult/Pediatric Airway Adapter Sets for 

infants/neonates as the airway adapter adds 6 ml dead space. 

WARNING:

 Do not apply negative pressure to remove condensed water from the NomoLine 

Family sampling line. 

WARNING:

 NomoLine ISA CO2 is not intended to be used for returning exhaust gases to the 

patient circuit. Exhaust gases should be returned to a scavenging system. 

Note:

 Disconnect the device from AC mains by removing the device cable connection from 

the host device. 

Содержание NomoLine ISA CO2

Страница 1: ...Operator s Manual NomoLine ISA CO2 Sidestream Gas Analyzer...

Страница 2: ......

Страница 3: ...ner inconsistent with the directions for use or labeling CAUTION Federal USA law restricts this device to sale by or on the order of a physician See instructions for use for full prescribing informati...

Страница 4: ......

Страница 5: ...nce Warnings and Cautions 11 Chapter 1 Description 13 General System Description 13 Features 13 NomoLine Sampling Lines 14 Chapter 2 Setting Up 15 System Setup 15 Chapter 3 Operation 17 LEGI Indicator...

Страница 6: ...Contents www masimo com 4 Masimo Repair Policy 30 Return Procedure 30 Sales End User License Agreement 30 Limited Warranty 31 Exclusions 31 Limitation of Warranty 31 End User License 32 No Implied Lic...

Страница 7: ...s cautions and notes A warning is given when actions may result in a serious outcome for example injury serious adverse effect death to the patient or user WARNING This is an example of a warning stat...

Страница 8: ......

Страница 9: ...d Use NomoLine ISA CO2 sidestream gas analyzer and accessories including NomoLine is intended to be connected to other medical backboard devices for monitoring of breath rate and the following breathi...

Страница 10: ......

Страница 11: ...ices with NomoLine ISA CO2 may result in damage to the device and or patient injury WARNING Carefully route the sampling line to reduce the risk of patient entanglement or strangulation WARNING Do not...

Страница 12: ...ecessary this equipment and the other equipment should be observed to verify that they are operating properly WARNING Portable and mobile RF communications equipment including peripherals such as ante...

Страница 13: ...y changes or modifications not expressly approved by Masimo shall void the warranty for this equipment and could void the user s authority to operate the equipment WARNING Use of accessories and cable...

Страница 14: ......

Страница 15: ...or a complete list of compatible NomoLine sampling lines cannulas and related consumables visit www masimo com Features Front View 1 NomoLine Capnography Input Connector Allows connection to NomoLine...

Страница 16: ...remains in a low power sleep mode Once the sampling line is connected the NomoLine ISA CO2 gas analyzer switches to measuring mode and starts delivering gas data For ordering information about NomoLin...

Страница 17: ...combination with N2O and or anesthetic agents 4 Connect the NomoLine ISA CO2 cable to the host device connection port 5 Connect a NomoLine sampling line to the NomoLine ISA CO2 input connector See Fr...

Страница 18: ......

Страница 19: ...escribed in the table LEGI Indicator Status Steady green light Capnography monitoring in operation and OK Blinking green light Zeroing in progress See Zeroing on page 30 Steady red light Sensor error...

Страница 20: ......

Страница 21: ...pling line not plugged in to NomoLine ISA CO2 completely Internal error Verify NomoLine ISA CO2 connection to the host monitor Disconnect and reconnect the sampling line to NomoLine ISA CO2 The NomoLi...

Страница 22: ...ampling line leaking Internal error Use the correct sampling line with NomoLine ISA CO2 Disconnect and reconnect the sampling line to NomoLine ISA CO2 Follow the Directions for Use to properly connect...

Страница 23: ...ll specified environmental conditions except for interference specified in Effects from Water Vapor Partial Pressure on Gas Readings on page 25 and Interfering Gas Vapor Effect on page 25 Gas Accuracy...

Страница 24: ...greater than 50 hPa in accordance to IEC 60601 1 12 CO2 measurement inaccuracy shall be 0 3 kPa 4 of reading Physical Characteristics Item Specification Dimensions 91 mm x 60 mm x 37 mm 3 6 x 2 4 x 1...

Страница 25: ...IPX4 Protection against splashing of water from all directions Mode of Operation Continuous operation Additional Specifications General Specifications Ambient CO2 800 ppm 0 08 vol Recovery Time After...

Страница 26: ...RR exceeds the RR threshold RRth according to the following formula for CO2 with NomoLine HH Adult Pediatric Airway Adapter Set Flags Breath Detected No Breath Detected Check Sampling Line Unspecified...

Страница 27: ...the NOMO section removes all condensed water no water will reach the NomoLine Capnography gas analyzer However at an ambient temperature of 37 C and a breathing gas with a relative humidity of 95 the...

Страница 28: ...0 vol Helium the actual measured CO2 concentration will typically be 1 0 06 5 0 vol 4 7 vol CO2 Note 4 According to the EN ISO 80601 2 55 standard Note 5 In addition to the EN ISO 80601 2 55 standard...

Страница 29: ...ge Fragile handle with care Keep dry Atmospheric pressure limitation Storage humidity limitation Do not use if package is damaged Gas inlet Date of manufacture YYYY MM DD Gas outlet Manufacturer Class...

Страница 30: ......

Страница 31: ...isopropyl alcohol Glutaraldehyde Solution Quaternary Ammonium Chloride Wipe 0 5 Sodium Hypochlorite Water Solution Accelerated Hydrogen Peroxide Maintenance Once every year it is recommended to perfor...

Страница 32: ...d enclose or include the following information and items or other information as the seller requires A letter describing in detail any difficulties experienced with the NomoLine ISA CO2 Warranty infor...

Страница 33: ...ed reassembled or repaired by anyone other than a person authorized by Masimo e used with other products like new sensors reprocessed sensors or other accessories not intended by Masimo to be used wit...

Страница 34: ...other device You may not disclose publish translate release distribute copies of modify adapt translate reverse engineer decompile disassemble or create derivative works based on the Masimo Product th...

Страница 35: ...licable Any use modification reproduction release performance display or disclosure of the Software including the related documentation by the U S Government or any of its agencies shall be governed s...

Страница 36: ......

Страница 37: ...e on Gas Readings 21 25 Electrical 22 End User License 32 Environmental 21 Exclusions 31 F Features 13 Front View 13 15 G General System Description 13 I Intended Use 7 Interfering Gas Vapor Effect 21...

Страница 38: ...oLine ISA CO2 Index www masimo com 36 Masimo Sales End User License Agreement 30 Sampling Line Replacement 14 19 20 Symbols 26 System Setup 13 15 T Troubleshooting NomoLine ISA CO2 19 Z Zeroing 17 25...

Страница 39: ......

Страница 40: ...www masimo com 300106 LAB 10152A 0219...

Отзывы: