User Manual
P a g e
|
40
of
52
Range:
1 to 9999
This parameter is enabled only for Periodic Trigger.
Logging
Time
(HH:MM:SS)
Total Logging Time in HH:MM:SS Format for Periodic Logging.
This parameter is enabled only for Periodic Trigger.
File No.
File Number.
Range:
1 to 25
This parameter is enabled only for Periodic Trigger.
File Info.
Shows the information of stored files. This information contains Logging Start
Time & Date and No of Samples stored in the file.
File Delete
Delete stored file.
START LOGGING
Press F1 key while selecting this button to start the Logging.
NOTES:
Maximum No of Reading that can be stored in,
Logging Mode
Max. Reading
Periodic
150000
Key Press
572
In Periodic mode, changing of any Measure or Source parameter
is not allowed. So While Periodic Logging is Running, User can’t
enter into
MEASURE
,
SOURCE
and
DISPLAY
menu. But in Key
Press Logging mode, there is no restriction.
In Periodic Mode, if error message
like “
Not Sufficient Memory
”
comes while starting the Logging. Try to Reduce Logging Time or
Increase Sampling Period or try deleting some existing files.
In Key Press Mode, If No of Samples reach its maximum limit that
is 484, the next sample will start from the first overwriting the
memory.
While Logging is running, entering into the
LOGGING
menu shows
below page.