User Manual
P a g e
|
38
of
52
Available Options:
Options
Icon
Description
None
-
No info is visible.
Bargraph
-
Shows the Horizontal bar Graph in 0% to
100% scale.
The value in Bar Graph depends on the
Display Mode and Source settings
Actual Value
Shows the Actual Output Value.
This option is not appear if
Main Display
is
Actual
.
This Parameter is enabled only for M Source or Switch test + Source
Display Mode.
4.3.3 Swtich Test Settings
This Page is appears in
RUN
MENU
DISPLAY
SWITCH TEST.
Parameter Name
Description / Options
Mode
Switch Test Operation Mode
Available Options:
2V(24Vdc,30mA)
Switch Close when External Switch (Potential
Free Contacts) short & Switch Open is External
Switch open.
Voltage Trigger
Switch Close when Input Voltage > Trigger
Voltage
Switch Open when Input Voltage < Trigger
Voltage
All the MEASURE Page parameters are disabled if Switch Test Display
Mode is selected.
Trigger Voltage
Trigger Voltage value for Voltage Trigger Switch Test Mode.
Range:
0 to 30 V
Enable only for Switch Test
Mode
as
Voltage Trigger.
Sound
Sound Setting for Switch Test Mode
Available Options:
Off
When Switch Close
Содержание LC-12