Ref. No: m94B/OM/101
Issue No: 00
Operator’s manual
Page 11 of 16
(8)
hh.mm.ss
(Time) Mode:
This mode shows the current hour, minute & second of data-logger. User can set hour,
minute & second via SMART-HT configuration software.
(9)
ltmr
(Log-Timer) Mode:
This mode shows recording interval of data-logger, user can change record interval via
menu key.
User can only configured
lTMR
value from below set:
{1,5,10,15,30,60,120,180,240,360,480,720,1440} this values are in minutes.
(10)
fdsp
(Fast Display) Mode:
When
fdsp on
sensor update data on LCD at every 15sec irrespective of
LTMR
value(record interval). when user START data-logging fast display status change to
fdsp off
and display update at interval of
LTMR
value(record interval).
(11)
mode
(Sensor Type) Mode:
This mode is for sensor selection. Sensor type of data-logger can be change with this
mode. User can change sensor type among this combination
t
(Only Temperature),
rh
(Only Humidity) &
trh
(Temp Humidity)
.
(12)
lmod
(Logging Mode) Mode:
Manual Start:
Device needs to be started by pressing Start/Stop button from SMART-HT configuration
software or by front key panel of data-logger.
Auto Start:
Device automatically starts recording on the specific date/time preset, no need to press
Start/Stop button after configuration.
(13)
min max
(Min-Max Reset) Mode:
This mode clears min-max value of temperature & humidity from data-logger. After
min-max reset data-logger start sorting new fresh min & max value from sensor.