Care and Maintenance

 

Calibration Check

 

It  is  rare  for  the  meter  calibration  to  change  or  the  electronics  to  malfunction 

without an obvious indication. If any ‘E’ codes display, return the instrument to 

the Martin Lishman service centre.

 

 

To verify the instrument calibration:

 

Use  the  Quickcheck  device  with  ATC  on  and  no  external  probe  connected. 

Ensure the Grain Cell is clean and dry. Set the instrument to Wheat. Place the 

Quickcheck on the Grain Cell and hold in position.

 

 

 

 
 

If the Grain Cell is not at 20°C:

 

For every whole degree above 20°C, ATC will subtract 0.1 %.  

 

Example: 

Quickcheck  reading  is  15.8%  at  a  cell  temperature  of  25°C.    ATC  subtracts  0.5% 

from 16.3% to give 15.8%.  Instrument is therefore calibrated and working correctly.

 

 

For every whole degree below 20°C, ATC will add 0.1%.

 

Example: 

Quickcheck  reading  is  16.6%  at  a  cell  temperature  of  17°C.    ATC  adds  0.3%  to 

16.3% to give 16.6%.  Instrument is therefore calibrated and working correctly.

 

 

Notes: 

 

 

If a calibration has a bias adjustment, Quickcheck readings will reflect this.  

 

 

 

Example: 

Quickcheck  reading  is  17.1%,  cell  temperature  is  17°C,  wheat  calibration  is 

biased  +0.5%.  The  reading  is  derived  from  16.3%  (Quickcheck  value  at  20°C)  +  0.5%  (bias)       

+ 0.3% (ATC) = 17.1%, so the meter is working correctly.

 

 

The  column  coloured  yellow  in  the  Calibration  Chart  on  Page  8  gives  the 

values for other crops with the Quickcheck at 20°C.  

 

 

Regular Servicing and Checks

 

According to most crop quality assurance schemes, your moisture meter should 

be checked annually  against oven  tested crop samples. It is also  important to 

ensure  that  vital  parts  such  as  the  grinder

-

compressor  are  working  correctly. 

The  Martin  Lishman  service  centre  is  fully  equipped  to  service  your  moisture 

meter  throughout  its  working  life.  We  also  conduct  testing  clinics  at  various 

locations around the country where you can check your meter and obtain a test 

certificate to satisfy your crop assurance scheme. 

 

See  our  website  for  the  latest  clinic  details  and  to  download  a  service  return 

form.

 

Press 

.

 

Take a reading.

 

While the reading is visible, press 

Λ

 or 

V,

 

the  display  will  then  show 

0.0

,  p

ress 

 

to display Grain Cell temperature 

 

The reading  is 

16.3% at 20°C, 

which is correct.  

 

Содержание Grainmaster i2-S

Страница 1: ...Martin Lishman Grainmaster i2 S Professional Crop Moisture Measurement Distributed by Martin Lishman Ltd Tel 44 0 1778 426600 www martinlishman com Instruction Manual GRN3000i2 S...

Страница 2: ...2 Other products in the Martin Lishman range See our website for further details www martinlishman com...

Страница 3: ...hip your moisture meter with the latest firmware but updates will become available You can update your meter with the use of the included USB cable and a Windows PC If you are using the Grainmaster mo...

Страница 4: ...usted manually Page 9 Display Bias Set Always use Automatic Temperature Correction ATC ATC adjusts readings if a sample is not at 20 C The default setting is ATC switched ON Page 10 Display flashing A...

Страница 5: ...Standby Mode Optional External Probe Socket USB Socket Meter Functions Items supplied Grain Hopper Yellow Black Pips Cleaning Brush Grinder Compressor Quick Check Grain Cup Grain Cell Spare Grinder B...

Страница 6: ...Ensure the Grain Cell is clean Align the Black Pips on the handle with the Yellow Slide the selector inwards to the Compress position Rotate handle anti clockwise until it clicks a few times Slide th...

Страница 7: ...roximately 7 seconds note and record this value as required To change the crop type Display Codes _ _ _ Standby mode U R Under measurement range The sample is too dry or too cold Can also display if t...

Страница 8: ...3 27 4 30 0 Sorghum Milo Ground 10 2 11 7 12 4 13 1 13 9 14 6 15 4 15 6 16 3 17 1 17 9 18 7 19 6 20 9 22 2 23 7 25 4 Soya Beans Ground 7 4 8 2 8 8 9 7 10 4 11 1 11 8 12 0 12 6 13 4 14 3 15 2 16 0 17...

Страница 9: ...ate for each crop and stored in the instrument s memory until removed as above Bias adjustment can also be made when using the Moisture Temperature or Bale Moisture Probes see Page 14 Temperature read...

Страница 10: ...rature Correction ATC corrects the moisture reading automatically Even with ATC on the grain sample and the meter Grain Cell need enough time to reach close temperature equilibrium If there are extrem...

Страница 11: ...meter version information will appear on screen before entering the setup menu The button Skip is used to navigate through the setup menu The and V buttons are used to change the selected parameter T...

Страница 12: ...e battery if not used for long periods When battery power is low displays A spare battery is included with the meter Separate the two halves of the unit by twisting the yellow locking ring anticlockwi...

Страница 13: ...t Quickcheck readings will reflect this Example Quickcheck reading is 17 1 cell temperature is 17 C wheat calibration is biased 0 5 The reading is derived from 16 3 Quickcheck value at 20 C 0 5 bias 0...

Страница 14: ...moisture probe reading Ensure the Grain Cell is clean Do not touch the cell while using the probe Push the probe into the grain or bale and allow a few minutes to stabilise Connect the probe to the me...

Страница 15: ...emperature probe reading Ensure the Grain Cell is clean Do not touch the cell while using the probe Push the probe into the grain or bale and allow the temperature to stabilise Connect the probe to th...

Страница 16: ...he cost of replacement or repair The manufacturers and Martin Lishman Ltd will not be liable for any consequential or indirect loss suffered by purchasers or users of the unit whether this loss arises...

Отзывы: