d) Character overrun
A general communications error occurred during the response
received from the loop device. Try the command again. If the error
persists, check the power and connections to the loop device.
e) Buffer overrun
A general communications error occurred during the response
received from the loop device. Try the command again. If the error
persists, check the power and connections to the loop device.
f) Framing bad
The structure of the response received from the loop device did not
match the HART protocol definition. Try the command again. If the
error persists, check the power and connections to the loop device.
g) Check byte bad
The response received from the loop device did not contain a valid
check byte per the HART protocol definition. Try the command
again. If the error persists, check the power and connections to the
loop device.
h) Invalid address
The response received from the loop device contained a HART
device address different from that sent. Ensure there are no burst
mode devices broadcasting on the loop. Try the command again. If
the error persists, check the power and connections to the loop
device.
i) Invalid command
The response received from the loop device contained a HART
command code different from that sent. Try the command again. If
the error persists, check the power and connections to the loop
device.
j) Invalid fixed data
The structure of the response received from the loop device did not
match the HART protocol definition. Try the command again. If the
error persists, check the power and connections to the loop device.