![Martel LC-110 Скачать руководство пользователя страница 20](http://html1.mh-extra.com/html/martel/lc-110/lc-110_reference-manual_1732377020.webp)
18
6.1.2 In Circuit, LC-110H Loop Power
In “mA Measure with 24V” mode the LC-110H is in circuit and the
loop power is supplied by the LC-110H as shown below.
(+)
(–)
MARTEL ELECTRONICS
LC-110H
Precision Loop
Calibrator w/HART™
100%
25%
0%
MENU
EXIT
Measure 24V
0.00%
4.000mA
250
Ω
Note:
In this situation the 250
Ω
HART resistor is enabled. If your loop
already has 250
Ω
, do not enable the HART resistor.
6.1.3 Across Circuit, Communicator Only
In “Communicator Only” mode the LC-110H is across the circuit and
the loop power is supplied externally as shown below.
MARTEL ELECTRONICS
LC-110H
Precision Loop
Calibrator w/HART™
100%
25%
0%
MENU
EXIT
PT12234
Device Setup & Data
Write LRV & URV Value
Trim, Set, & Zero
Device Diagnostic
Data Log & Config Log
24V
SUPPLY
(+)
(–)
4 to 20mA
RL≥250Ω
Note:
In the Communicator Only Mode, there must be 250
Ω
s
resistance present in the loop.
www.
GlobalTestSupply
.com
Find Quality Products Online at: