6.12 Error Displays
There are two classes of errors which may occur during HART
communications. The first class contains errors that are reported by
the loop device in its response. The second class contains errors
that are detected by the LC110H when receiving responses from the
loop device.
Loop Device Errors
These errors are reported by the loop device in its response to the
LC110H, per the HART protocol.
They are indicated by the text “Loop Device error:” at the top of the
display. This is followed by one of the following error messages.
1) Invalid selection
The data value sent with the HART command is not allowed by this
loop device. The command was ignored by the loop device. HART
response code 2.
2) Value too large
The data value sent with the HART command is greater than the
range allowed by this loop device. The command was ignored by
the loop device. HART response code 3.
3) Value too small
The data value sent with the HART command is less than the range
allowed by this loop device. The command was ignored by the loop
device. HART response code 4.
4) Message too short
The HART command received contains fewer characters than
expected by this loop device. The command was ignored by the
loop device. HART response code 5.
5) Device specific error
The loop device has detected a device specific error. The LC110H
does not support the reporting of details on device specific errors.
The command was ignored by the loop device. HART response
code 6.