12.8 Thermocouple Specification, Output and Input
Absolute Uncertainty,
tcal ±5 °C, ±(°C)
1,2
Range (°C)
Output/Input
TC Type
Minimum
Maximum
90 days
1 year
B
600 °C
800 °C
0.42 °C
0.46 °C
800 °C
1550 °C
0.39 °C
0.39 °C
1550 °C
1820 °C
0.44 °C
0.45 °C
C
0 °C
150 °C
0.25 °C
0.30 °C
150 °C
650 °C
0.21 °C
0.26 °C
650 °C
1000 °C
0.23 °C
0.31 °C
1000 °C
1800 °C
0.38 °C
0.50 °C
1800 °C
2316 °C
0.63 °C
0.84 °C
E
-270 °C
-100 °C
0.38 °C
0.50 °C
-100 °C
-25 °C
0.16 °C
0.18 °C
-25 °C
650 °C
0.14 °C
0.16 °C
650 °C
1000 °C
0.16 °C
0.21 °C
J
-210 °C
-100 °C
0.20 °C
0.27 °C
-100 °C
-30 °C
0.18 °C
0.20 °C
-30 °C
760 °C
0.14 °C
0.17 °C
760 °C
1200 °C
0.18 °C
0.23 °C
K
-270 °C
-100 °C
0.25 °C
0.33 °C
-100 °C
-25 °C
0.19 °C
0.22 °C
-25 °C
120 °C
0.14 °C
0.16 °C
120 °C
1000 °C
0.19 °C
0.26 °C
1000 °C
1372 °C
0.30 °C
0.40 °C
L
-200 °C
-100 °C
0.37 °C
0.37 °C
-100 °C
800 °C
0.26 °C
0.26 °C
800 °C
900 °C
0.17 °C
0.17 °C
N
-270 °C
-100 °C
0.33 °C
0.40 °C
-100 °C
-25 °C
0.20 °C
0.24 °C
-25 °C
410 °C
0.16 °C
0.19 °C
410 °C
1300 °C
0.21 °C
0.27 °C
R
-50 °C
250 °C
0.58 °C
0.58 °C
250 °C
400 °C
0.34 °C
0.35 °C
400 °C
1000 °C
0.31 °C
0.33 °C
1000 °C
1767 °C
0.30 °C
0.40 °C
S
-50 °C
250 °C
0.56 °C
0.56 °C
250 °C
1000 °C
0.36 °C
0.36 °C
1000 °C
1400 °C
0.30 °C
0.37 °C
1400 °C
1767 °C
0.35 °C
0.46 °C
T
-270 °C
-150 °C
0.51 °C
0.63 °C
-150 °C
0 °C
0.18 °C
0.24 °C
0 °C
400 °C
0.13 °C
0.16 °C
U
-200 °C
0 °C
0.56 °C
0.56 °C
0 °C
600 °C
0.27 °C
0.27 °C