HART_OFF
IEEE-488
RS-232
Sequential
Overlapped
This command disables the isolated DC current input HART resistor.
Parameter:
<None>
Response:
<None>
Example:
HART_OFF
This example disables the isolated DC current input HART resistor.
HART_ON
IEEE-488
RS-232
Sequential
Overlapped
This command enables the isolated DC current input HART resistor.
Parameter:
<None>
Response:
<None>
Example:
HART_ON
This example enables the isolated DC current input HART resistor.
*IDN?
IEEE-488
RS-232
Sequential
Overlapped
This command returns the unit identification: instrument manufacturer, model number,
serial number, and firmware revision level.
Parameter:
<None>
Response:
<text string>
where <text string> contains the following four fields separated by
commas:
1. Manufacturer
2. Model number
3. Serial number (always 0)
4. Firmware revision level
Example:
*IDN?
MARTEL, 3001,0,1.2
This example indicates the manufacturer is Martel, the model is 3001,
the serial number is 0, and the firmware version is 1.2.