5. Isolated Inputs
5.1 Voltage Input
The 3001 can measure DC voltages from 0 V to 100 V, using the following two ranges for
maximum accuracy: 10 V, and 100 V.
a) Disconnect any test leads from external devices.
b) Press the
key to select isolated DC voltage and current input mode, if not
already selected. If the desired DC voltage mode is not displayed, press the key
again to cycle to the desired DC voltage mode.
c) Connect the unit under test to the isolated voltage/current input terminals of the 3001
as shown in figure 17.
Figure 17 - Isolated DC Voltage Input Connection
5.2 Current Input
The 3001 can measure DC current from 0 mA to 50 mA.
a) Disconnect any test leads from external devices.
b) Press the
key to select isolated DC voltage and current input mode, if not
already selected. If the DC current mode is not displayed, press the key again to
cycle to it.
c) Connect the unit under test to the isolated voltage/current input terminals of the 3001
as shown in figure 18.