OPER?
IEEE-488
RS-232
Sequential
Overlapped
This command returns the present operate/standby mode setting.
Parameter:
<None>
Response:
<value>
where <value> is "1" for operate mode and "0" for standby mode
Example:
OPER?
1
This example indicates that the 3001 is in operate mode.
*OPT?
IEEE-488
RS-232
Sequential
Overlapped
This command returns a list of the installed hardware and software options. This
command is reserved for future use.
Parameter:
<None>
Response:
<text string>
where <text string> is "0" if no options are installed, or a list of installed
options separated by commas
Example:
*OPT?
0
The example indicates that no options are installed.
OUT
IEEE-488
RS-232
Sequential
Overlapped
This command sets the output mode and value of the 3001. To source a temperature,
select the desired mode and sensor parameters first with the TSENS_TYPE, RTD_TYPE,
and TC_TYPE commands. Use the multiplier prefixes k for kilo, m for milli, and u for
micro with the OUT command units, as desired. The unit may be omitted if the output
mode is not to be changed.
Parameter:
<value> <units>
where <value> is the output value to be generated
and where <units> is one of the following, with a possible prefix
multiplier (k, m, or u) as described above:
V
DC volts
A
DC current
OHM
Resistance
CEL
Temperature in Celsius
FAR
Temperature in fahrenheit
Response:
<None>
X
X
X
X
X
X
X
X
X