3. Operation

 

 
3.1 Keyboard and display

 

 

The Mark 25 front panel consists of a    
4 digit red LED display, 3 parameter and 
channel LED’s, an alarm LED, and an 8-
key keypad. The display is used to 
present all parameter measurements, 
operator prompts and system errors. 
 
 
 
 
 
 
 
 
 
 
 
 
 
The keys are arranged in two rows, four 
keys in each row.  
 
The 

top row

 

of keys

 consist of three 

color-outlined 

function keys

 and are 

used to perform instrument operations 
such as selecting which parameter to 
view, what range the parameter is to be 
viewed in, parameter calibration and 
system set up. 
 
The 

bottom row of keys

 are multi-

colored, multi-legend keys and are used 
in conjunction with the function keys. 
The same colored legends correspond 
to the same colored function key. 
 
The 

arrow keys

 are used to view 

channels, select functions during setup 
and calibration, and increment or 
decrement a displayed value. 
 

 
 
READ key

 - When the <READ> key is 

pressed, the Mark 25 will display 

rd

. If 

one of the bottom keys is selected, the 
parameter high-lighted in red will be 
displayed and its corresponding LED lit. 

 

 
 
RANGE key

 - Range changes can be 

made manually for raw conductivity and 
temperature-corrected conductivity only. 
To change the range, press the 
<RANGE> key, then select 

 

either LOW, MED, or HIGH to display 
the desired range.  If the Mark 25 is 
setup for auto-ranging, the Mark 25 will 
display 

AUrg

 as long as the <RANGE> 

key is pressed. 
 

NOTE: 

Manually changing the range for 

temperature-corrected conductivity 
changes the analog recorder output 
scaling.

 

See Section 3.7 for details

.

 

 
Viewing a different channel

 is easily 

done by pressing the <UP> or <DOWN> 
arrow key.  The parameter LED’s to the 
left of the Mark 25 display also indicate 
which channel is being viewed. Channel 
ID’s are as follows: 
 

Red 

LED - Channel 1 

Green

 LED - Channel 2 

Red/Green 

LED - Channel 3 

Flashing Red 

LED - Channel 4

 

 
 
3.2 The CALIB key

 

 
All setup and calibration functions are 
conducted via the <CALIB> key. The 
<CALIB> allows you to set the following: 
 

READ

 

Range

 

CALIB

 

 

LOW 

o

ZERO

 

MED 
RAW 

SPAN

 

HIGH 

TC 

RUN

 

 

Содержание Mark 25

Страница 1: ...Martek Instruments Inc Mark 25 Ultrapure Water Conductivity Analyzer Instruction Manual...

Страница 2: ...rial number of the product and a brief description of the problem All shipments to MARTEK must be freight prepaid and addressed as follows MARTEK INSTRUMENTS INC 2609 Discovery Drive Raleigh NC 27616...

Страница 3: ...ectrical connections to the monitor 6 3 Operation 3 1 Keyboard display 7 3 2 The CALIB Key 7 3 3 Calibrating Temperature 8 3 4 Calibrating Conductivity 9 3 5 Setting Alarms 9 3 6 Setting Recorder Outp...

Страница 4: ...r outputs 0 20 mA and a serial RS232 digital output for hookup to external recording equipment For maximum flexibility alarm set points and analog outputs can be assigned to any range on any parameter...

Страница 5: ...utput Alarm Output Four solid state relays 5 amp 250 V Weight 2 lbs 0 9 kg including one sensor with 10 feet 3 meters Construction Readout Module Case ABS plastic NEMA 12 IP65 Sensor Type 316 Stainles...

Страница 6: ...Installation 2 1 Panel Cutout for Monitor 3 60 3 60 Panel Cut out 2 2 Sensor Dimensions 2 75 6 98 cm 1 75 4 44 cm Side View 8 00 20 32 cm 1 12 2 84 cm Top View 125 31 cm 196 49 cm 2 places 4 125 10 47...

Страница 7: ...erly grounded outlet If recorder digital or alarm outputs are to be used make certain the Mark 25 is unplugged before connecting the cable to the appropriate recorder or computer Follow the wiring dia...

Страница 8: ...ssed the Mark 25 will display rd If one of the bottom keys is selected the parameter high lighted in red will be displayed and its corresponding LED lit RANGE key Range changes can be made manually fo...

Страница 9: ...rk 25 should be calibrated to ASTM or NIST traceable standards To calibrate temperature press the CALIB key The display will flash o C To select temperature press the UP ARROW key The flashing will st...

Страница 10: ...y channel Press the CALIB until the display indicates the alarm prompt ALA Next press the UP ARROW key The display will prompt with AL 1 and all parameter LED s will be lit The UP ARROW key can be use...

Страница 11: ...nction is ON the decimal point can be moved by pressing the RANGE key Press the CALIB key to select the desired parameter At this point the standard 4 20 mA output range can be changed to any value de...

Страница 12: ...esired value has been reached press the CALIB key to set and activate the password Once the password is enabled the display will default to 1999 whenever the password prompt is selected To gain access...

Страница 13: ...will be lit and flashing Use the UP or DOWN ARROW key to advance to the auto ranging prompt AUrG When the CALIB key is pressed the Mark 25 will display OFF To activate the auto ranging function press...

Отзывы: