4
1.2 Specifications
Parameter
Range
Accuracy
Resolution
Conductivity
0-1.000
µ
S/cm
+0.01
µ
S/cm
0.001
µ
S/cm
0-10.00
µ
S/cm
+0.1
µ
S/cm
0.01
µ
S/cm
0-100
µ
S/cm
+1.0
µ
S/cm
0.1
µ
S/cm
Temperature
0-100
o
C
+0.1
o
C
0.01
o
C
Power:
97-240 VAC, 50/60 Hz, 24Vdc, fuse protected
Recorder Output:
Four isolated, scaleable 0-20 mA DC analog & RS232C serial ASCII digital output
Alarm Output:
Four solid-state relays, 5 amp, 250 V
Weight:
2 lbs (0.9 kg) including one sensor with 10 feet (3 meters)
Construction:
Readout Module Case: ABS plastic, NEMA 12 (IP65)
Sensor: Type 316 Stainless Steel and Teflon
Cabling:
8-conductor, 24 gauge, PVC jacket, shielded, 0.25 inch (0.63 cm)
Dimensions:
¼ DIN case, 6.51 x 3.78 x 3.78 (165.5 x 96 x 96 mm) LWH
Installation:
Panel mount (mounting hardware included). Recommend mounting sensor
vertically with inlet at the bottom. Position of sensor cabling not a factor in
mounting.
Flow rate:
45-250 ml/min