13
on the front panel, these settings will be ignored, except for
Auto Return
or
Cycling
. If either of
these functions is turned on, PC control will be turned off.
Default setting: CONSOLE
Available settings: CONSOLE, PC
Label
Description
+ or –
Cycles through the available settings
ENTR
Returns to the Function Setup menu
ESC
Exits the function without saving changes
4.8 Baud rate
This setting corresponds to the baud rate setting of the computer program controlling the test stand.
Default setting: 19200
Available settings: 1200, 2400, 4800, 9600, 19200, 38400, 57600, 115200
Label
Description
+ or –
Cycles through the available settings
ENTR
Returns to the Function Setup menu
ESC
Exits the function without saving changes
4.9 Stop bits and parity (STOP & PAR)
This corresponds to the stop bits and parity settings of the computer program controlling the stand.
Default setting: 8-1n
Available settings:
8-1E
8 stop bits, 1 stop bit, even parity
8-1o
8 stop bits, 1 stop bit, odd parity
8-1n
8 data bits, 1 stop bit, no parity
8-2n
8 data bits, 2 stop bits, no parity
7-1E
7 data bits, 1 stop bit, even parity
7-1o
7 data bits, 1 stop bit, odd parity
7-2E
7 data bits, 2 stop bits, even parity
7-2o
7 data bits, 2 stop bits, odd parity
7-2n
7 data bits, 2 stop bits, no parity
C O N T R O
L : C O N S O L E
E S C − + E N T R
B A U D R A T E : 1 1 5 2 0 0
E S C − + E N T R
S T O P & P A R : 8 - 1 n
E S C − + E N T R
Содержание TSTM-DC
Страница 1: ...Series TSTM DC Motorized Torque Test Stands User s Guide...
Страница 23: ...Series TSTM DC Test Stands User s Guide 23 10 DIMENSIONS in mm TSTM DC...
Страница 24: ...Series TSTM DC Test Stands User s Guide 24 TSTMH DC CONTROL PANEL...
Страница 25: ...Series TSTM DC Test Stands User s Guide 25 NOTES...