Mark-10 TSTM-DC Скачать руководство пользователя страница 1

 

Series

 

TSTM-DC 

 

Motorized Torque Test Stands 

 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

User’s Guide 

 
 
 
 

 
 
 
 
 

 

Содержание TSTM-DC

Страница 1: ...Series TSTM DC Motorized Torque Test Stands User s Guide...

Страница 2: ...p safety and operation instructions Dimensions and specifications are also provided For additional information or answers to your questions please do not hesitate to contact us Our technical support a...

Страница 3: ...g the pair of screws at each end Install the indicator mounting kit on the column For the vertical models install the base with four 5 16 18 x 1 screws and the top cap with four ribbed push in rivets...

Страница 4: ...the control panel can be mounted at any angle although extra care should be taken during installation and operation IMPORTANT Do not fasten any screws more than 0 25 in 6 mm into the base of the cont...

Страница 5: ...lows for PC control if appropriate functions are enabled Plug one end of the 09 1056 serial cable into this connector and the other end into a serial connector on a computer or USB converter 4 Power P...

Страница 6: ...ances or products items that can shatter in an unsafe manner and any other components that can present an exceedingly hazardous situation when acted upon by torque When using a grip ensure that it sec...

Страница 7: ...worn if a destructive failure of a test sample is possible 3 Keep away from moving parts of the test stand Loose articles of clothing should not be worn Long hair should be covered to avoid a hazardo...

Страница 8: ...cycles password editing and other functions 2 OPERATING MODE This is the operating mode in which testing sequences can be started and stopped SOFT KEYS Functions are determined by the corresponding t...

Страница 9: ...of settings before saving 4 1 Speed Independent speeds can be set for both directions of movement Default setting 10 RPM 60 s Available settings TSTM DC TSTMH DC 0 01 15 RPM 0 05 90 s TSTM DC 2 TSTMH...

Страница 10: ...mpletions in the M7I whichever occurs first at the specified speed stopping for the specified amount of dwell time and returning to the other limit at the specified speed Default setting 00000 off Ava...

Страница 11: ...he next number will be 9999 9 and continue decrementing ENTR Returns to the Test Function Setup menu ESC Exits the function without saving changes 4 5 CW and CCW travel limits CW L and CCW L This sett...

Страница 12: ...D and CCW OVERLOAD This setting corresponds to the percentage of full scale at which torque plate rotational travel stops For example a setting of 80 for a 100 lbFin capacity torque sensor would stop...

Страница 13: ...400 57600 115200 Label Description or Cycles through the available settings ENTR Returns to the Function Setup menu ESC Exits the function without saving changes 4 9 Stop bits and parity STOP PAR This...

Страница 14: ...tton has been pressed and held Subsequently pressing STOP will stop torque plate motion 2 Momentary The torque plate will rotate only if the button is held down Releasing the button will stop movement...

Страница 15: ...indicates that the password is disabled and that the user can freely enter the Test Function Setup menu Default setting 0000 off Available settings 0000 9999 Label Description or Increases or decreas...

Страница 16: ...ses and moves at maximum speed to the other limit whichever occurs first 3 Cycling Mode Test stand cycles between limits at the selected speed and pauses at each limit for a selected period of time 4...

Страница 17: ...in motion an LED indicator on the button pushed will be illuminated The KEYS setting controls how test stand movement responds to the push of either of the directional buttons The three settings are 1...

Страница 18: ...the set speed to whichever limit it encounters first and stops Then it returns at maximum speed to the opposite limit whichever occurs first and stops Press either directional button to initiate an Au...

Страница 19: ...equence Press either directional button to resume Once the sequence has been completed the screen will revert to the number of cycles programmed originally To begin another cycle test press either dir...

Страница 20: ...quired g 000007 5 7 5 Deg g000010 7 10 7 Deg h Set CW travel limit Revolutions format h XXXX XX Negative sign if applicable four leading zeroes and two decimal places required h 0010 55 10 55 Turns h0...

Страница 21: ...the test stand 2 Attempt to loosen subcomponents of the test stand ex fasteners brackets etc All components should be firmly attached If any looseness is detected stop using the test stand and contact...

Страница 22: ...ed variation with load 0 Stepper motor driven Speed accuracy 0 2 of setting Rotational travel accuracy 0 7 0 02 rev Rotational travel repeatability 0 1 0 01 rev Rotational travel resolution 0 1 0 01 r...

Страница 23: ...Series TSTM DC Test Stands User s Guide 23 10 DIMENSIONS in mm TSTM DC...

Страница 24: ...Series TSTM DC Test Stands User s Guide 24 TSTMH DC CONTROL PANEL...

Страница 25: ...Series TSTM DC Test Stands User s Guide 25 NOTES...

Страница 26: ...pport In addition to our standard line of products we can provide modifications and custom designs for OEM applications Our engineering team is eager to satisfy any special requirements Please contact...

Отзывы: