Page 1 of 23
11/14/2006
Rev. 14
NOTE:
Please specify the following parameters when ordering a transmitter; process type, process
range (%, ppm), thermocouple type, temperature scale F/C, analog output 1 process and
scale, analog output 2 process and scale.
Typical Oxygen Transmitter Calibration
(F840030)
Calibration
Function
Measured Value or
Input
Output / Units
Cold Junction
Room Temp
°F
Thermocouple
min
800°F (B type)
standard t/c type
°F
Thermocouple
max
3000°F (B type)
standard t/c type
°F
Millivolt
0.0 mV
Millivolts
Millivolt
2000 mV
Millivolts
Analog 1 Zero
0% O2
4.0 mA +/- 0.1
Analog 1 Span
20.9% O2
20.0 mA +/- 0.1
Analog 2 Zero
800°F +/- 5°
4.0 mA +/- 0.1
Analog 2 Span
3000°F +/- 5°
20.0 mA +/- 0.1
Typical Carbon Transmitter Calibration
(F840031)
Calibration
Function
Measured Value or
Input
Output / Units
Cold Junction
Room Temp
°F
Thermocouple
Min
MUST BE
SPECIFIED
°F
Thermocouple
Max
MUST BE
SPECIFIED
°F
Millivolt
0.0 mV
Millivolts
Millivolt
2000 mV
Millivolts
Analog 1 Zero
0% Carbon
4.0 mA +/- 0.1
Analog 1 Span
2.55% Carbon
20.0 mA +/- 0.1
Analog 2 Zero
MUST BE
SPECIFIED
4.0 mA +/- 0.1
Analog 2 Span
MUST BE
SPECIFIED
20.0 mA +/- 0.1