Page 16 of 23
11/14/2006
Rev. 14
BLOCK 0
HEX
DEC
PARAMETER
DESCRIPTION
READ/WRITE
source value is based on the selection in
ASRC upper byte where
14
20
SPARE
SPARE
READ/WRITE
15
21
SPARE
SPARE
READ/WRITE
16
22
SPARE
SPARE
READ/WRITE
17
23
TEMPFIL
Temperature Input Filter in seconds
Range = 0 to 3276. The higher the number
the faster the reading update.
DEFAULT = 1000
READ/WRITE
BLOCK 1
HEX
DEC
PARAMETER
DESCRIPTION
READ/WRITE
18
24
MVFIL
Millivolt Input Filter in seconds
Range = 0 to 3276. The higher the number
the faster the reading update.
DEFAULT = 1000
READ/WRITE
19
25
AZERO
LINEAR OFFSET, Y INTERCEPT LINEAR
SCALING FOR INPUT A
READ/WRITE
1A
26
ANUM
LINEAR SPAN VALUE FOR INPUT A
READ/WRITE
1B
27
BZERO
LINEAR OFFSET, Y INTERCEPT LINEAR
SCALING FOR INPUT B
READ/WRITE
1C
14
BNUM
LINEAR SPAN VALUE FOR INPUT B
READ/WRITE
1D
15
PROC
This value is the calculated process value
shown as an integer. The decimal point and
exponent values are required to determine the
actual scaled value.
Range = -999 to 9999.
For example: If the process = oxygen, display
decimal point = 2, and exponent = 6, and
PROC = 1234, then the actual value and
displayed as 12.34 ppm.
READ ONLY
1E
16
COLDJCT
COLD JUNCTION
Where 1 COUNT = 1°F (°C), RANGE = -99 TO
255°F (°C). Note this parameter is an
unsigned integer.
READ ONLY
1F
17
TEMP
MEASURED TEMPERATURE
Where temperature is presented in degrees C
or F, based on the C/F setting. Note this
parameter is an unsigned integer of
temperature -2721 = 62815
Range = max / min range of selected
thermocouple.
READ ONLY
20
18
MV
MEASURED MILLIVOLT
Where this value is scaled in 0.1 mV
increments, i.e. 10001 = 1000.1.
Range = 0 to 2000 mV.
READ ONLY
21
19
DACV1
ANALOG OUTPUT 1
0 to 4095 is 4 to 20 mA In dual mode 4mA = -
100, 12mA = 0, 20mA = +100
READ/WRITE