Table 1 Process Parameters

Parameter Name

Selection

Default

Units or Options

Range

PROCESS TYPE

%O2

CARBON, DPT,
%O2, MV

CARB PROC FACT

150

0 to 1000

DEWPT PROC FACT

150

0 to 1000

OXYGEN EXPON

0002

POWER OF TEN

0 to 31

TC TYPE

B

B, C, E, J, K, N,
NNM, R, S, T

Process Type

Selecting the process type determines what type of calculation the Smart Transmitter is
going to do based on the probe millivolt and probe temperature inputs. The default process
value for the Smart Transmitter is %O2 with an exponent selection of 2. This is the
selection most often used in Boiler control and Combustion applications.

Percent Carbon and dew point are typically processes that are used in steel treating
applications. Percent Carbon is the process value most often used for the control of case
depth or the percent of carbon in a steel hardening furnace. Dew Point is used in the control
for endothermic generators.

Carbon Process Factor

The carbon process factor can be used to adjust the % carbon value. This number takes
into account a number of assumptions that the carbon value is based on. Primary among
these is the assumed level of CO in the atmosphere. See the Theory of Process Calculation
section for a complete explanation of this value.

It maybe necessary to change the apparent furnace carbon as measured by the oxygen
probe if this value is different than actual load samples, shim stocks, or gas analysis. The
basic rule of thumb is that an increase is the carbon process factor will decrease the
apparent carbon level in the furnace. The default value is 150. Typical values can very
from 50 to 400. Increase or decrease the process factor until the desired carbon level is
achieved. A process factor that is drastically different than normal may be an indication of
a failing probe, water or air leak in the furnace, or excess methane present. Refer to probe
troubleshooting guides to determine what other factors maybe effecting the carbon value.

Dew Point Process Factor

The dew point process factor is similar to the carbon process factor but is used to adjust the
dew point value if dew point is selected as the process value. This number takes into
account a number of assumptions that the dew point value is based on. Primary among
these is the assumed level of hydrogen in the atmosphere. See the Theory of Process
Calculation section for a complete explanation of this value.

Содержание Oxymit F200060

Страница 1: ...Marathon Sensors Inc Oxymit Transmitter Operators Manual...

Страница 2: ...lication are duly marked and the sole property of their respective owners No attempt at trademark or copyright infringement is intended or implied Marathon Sensors makes no warranties express or impli...

Страница 3: ...5 Carbon Process Factor 5 Dew Point Process Factor 5 Oxygen Exponent 6 TC Type 6 ANALOG OUTPUT CHANNELS 6 CALIBRATION 7 PROCESS VARIABLE CALCULATIONS 8 PERCENT OXYGEN 8 PERCENT CARBON 8 DEWPOINT 8 COM...

Страница 4: ...000 F B type standard t c type F Millivolt 0 0 mV Millivolts Millivolt 2000 mV Millivolts Analog 1 Zero 0 O2 4 0 mA 0 1 Analog 1 Span 20 9 O2 20 0 mA 0 1 Analog 2 Zero 800 F 5 4 0 mA 0 1 Analog 2 Span...

Страница 5: ...ent stand alone sensor The transmitter is located near the probe preferably mounted in an enclosure The transmitter mounts onto a DIN rail and requires a 24VDC power supply It measures the probe tempe...

Страница 6: ...load of 0 5 amps DC The Probe Transmitter is not rated to act as a safety device It should not be used to provide interlocking safety functions for any temperature or process functions Alarm capabilit...

Страница 7: ...gh current power cables Where it is impractical to do this use shielded cables with the shield grounded at the Probe Transmitter enclosure ground as show above Parameter Selections The following table...

Страница 8: ...n Primary among these is the assumed level of CO in the atmosphere See the Theory of Process Calculation section for a complete explanation of this value It maybe necessary to change the apparent furn...

Страница 9: ...60 NNM 32 0 2000 1090 R 300 150 3000 1650 S 300 150 3000 1650 T 32 0 700 370 The Cold Junction correction is applied to all thermocouple types Analog Output Channels The analog outputs are factory con...

Страница 10: ...999 9999 PROG 0 4095 OUTPUT 2 MODE TEMP 800 3000 F 4 20mA O2 CARBON DEWPT TEMP LIN PROG O2 0 9999 C 0 00 2 55 DP 99 9 212 0 Temp 999 3000 LIN 999 9999 PROG 0 4095 NOTE SEE PAGE 4 FOR TYPICAL CALIBRATI...

Страница 11: ...0215 Tk Where E probe millivolts Tk probe temperature in degrees Kelvin The 20 95 is the O2 in air Percent Carbon e E 786 0 043102 Tk C 5 102 29 PF 400 e E 786 0 043102 Tk Where E probe millivolts Tk...

Страница 12: ...to these values can be made by writing to the appropriate memory register The Transmitter communicates in Modbus RTU Remote Terminal Unit protocol using 8 bit binary data characters Message character...

Страница 13: ...Transmitter is concerned they are all just memory locations but the response to each command is consistent with Modbus specifications The high order bit in this field is set by the slave device to ind...

Страница 14: ...ucture The following is an example of a function 06 call to change data in register 01 to 200 The response from the instrument confirms the new value as being set Transmit from Host or Master Address...

Страница 15: ...S 8 9 PARITY SETTING 00 Even Parity 7 bits 1 Stop bit 01 No Parity 8 bits 1 Stop bit 10 Odd Parity 7 bits 1 Stop bit BITS 10 11 RESPONSE DELAY 0 No delay applied to response 1 10ms delay applied to re...

Страница 16: ...LD JUNCTION TRIM COLD JUNCTION TRIM unsigned integer RANGE 128 TO 127 WHERE 1 COUNT 1 DEG C or F and 128 65408 HIGH BYTE HOST ADDRESS BITS 0 7 RANGE 0 255 07 7 SPARE SPARE 08 8 CONFIG0 Input Configura...

Страница 17: ...Fault BITS 6 7 SPARE BIT 8 CPU Fault BIT 9 Min Idle counter 0 BIT 10 Keyboard failure stuck key or a key was pressed during power up BIT 11 Flash Erase Failed BIT 12 Flash Checksum Failed BIT 13 EEPRO...

Страница 18: ...0 to 100 PO 0C 12 DAC_OFFSET_1 DAC 1 OFFSET CALIBRATION READ WRITE 0D 13 DAC_SPAN_1 DAC 1 SPAN CALIBRATION READ WRITE 0E 14 DAC_OFFSET_2 DAC2 OFFSET CALIBRATION READ WRITE 0F 15 DAC_SPAN_2 DAC2 SPAN...

Страница 19: ...WRITE 1C 14 BNUM LINEAR SPAN VALUE FOR INPUT B READ WRITE 1D 15 PROC This value is the calculated process value shown as an integer The decimal point and exponent values are required to determine the...

Страница 20: ...2 0 to 4095 is 4 to 20 ma In dual mode 4mA 100 12mA 0 20mA 100 READ WRITE 23 35 SPARE SPARE 24 36 SPARE SPARE 25 37 SPARE SPARE 26 38 SPARE SPARE 27 39 SPARE SPARE 28 40 SPARE SPARE 29 41 SPARE SPARE...

Страница 21: ...llivolts 0 1 millivolt Input Impedance 25 Megohm Cold junction compensation 1 deg F DC outputs Isolated 0 to 20mA 650 max Isolation 1000V DC AC Power input to signal inputs Power input to communicatio...

Страница 22: ...s Wire Size AWG 24 12 flexible stranded removable terminal blocks Max Torque 0 8 Nm CAUTION DO NOT CONNECT OR DISCONNECT HOUSING PLUGS WHILE MODULE IS POWERED OR UNDER LOAD Weight 10 oz Environmental...

Страница 23: ...1 2 RF Interference EN 55011 Enclosure class A Power main class A Note This instrument is designed for installation inside a grounded metal enclosure Always observe anti static precautions when instal...

Отзывы: