Initial operation
Assembly and connection of accessories
16
GA 5750.4172 EN 14
3
3.4
Assembly and connection of accessories
DANGER!
Danger to life!
Hazard resulting from incorrect handling.
Be absolutely sure to observe the operating instructions for all the products used in
the configuration.
WARNING!
Configuration of accessory!
The design of the accessory is crucial for the proper function of the tapping unit.
Use of improper accessory will reduce the tapping unit's performance or even
cause the unit to fail.
All devices and accessories which are combined with the tapping unit must be
listed in the accessories list or meet the specifications of the interface description.
WARNING!
Tensile forces!
Hold the basic unit with one hand when installing or removing accessories in order
to compensate for the tensile forces which are created.
NOTE
Connection thread of hand-held nebuliser: M 15x1
Connection thread for O
2
connection tube to regulating unit: G 3/8“
Connection thread for Air connection tube to regulating unit: M 18x1
3.4.1
Connection of FINA FLOW O 15
2
1
3
4
5
6
Fig. 3: Connection of FINA FLOW O 15
Ensure that the coupling ring of the
connection tube is fitted with flat gaskets.
Ensure that the adapter (1) contains a flat
gasket.
Screw adapter directly to the flowmeter
output (2) and tighten fingertight until stop.
Tighten the connection tube (3) with the
larger coupling ring (4) fingertight onto the
connector.
Tighten the connection tube with the
smaller coupling ring (5) onto the connector
(6) of the hand-held nebulizer.
Pay attention that no tensile forces affect
the hand-held nebulizer.