Introduction
Incorrect operation/contraindications
10
GA 5750.4172 EN 14
1
1.5.4.1
Incorrect operation/contraindications
The hand-held nebuliser is not approved for the following purpose or under the following
conditions:
• for oil-based medication,
• for medication generating an explosive mixture with oxygen,
• for connection to respiratory systems,
• in MR areas,
• for application in intubated patients,
• • for other medical gases except O2 and Air.
1.5.4.2
Possible applications
Connection to O2 flowmeter:
The 9/16“ - G 3/8“ adapter and the O2 ISO M 15x1 - G 3/8“ connection tube are used for
connection to an O2 flowmeter. The flow of volume supplied to the hand-held nebuliser is
regulated by:
• S FLOW O 15
• S DFLOW O 15
• LS FLOW O 15
• LS DFLOW O 15
• FINA FLOW O 15
• FINA DFLOW O 15
Connection to Air flowmeter:
The 9/16“ - M 18x1 adapter and the Air M 15x1 - M 18x1 FN / ISO connection tube are used for
connection to an Air flowmeter. The flow of volume supplied to the hand-held nebuliser is
regulated by:
• S FLOW A 15
• LS FLOW A 15
• FINA FLOW A 15
Connection to FINA O2 precision regulation valve
• The flow of volume supplied to the hand-held nebuliser is regulated using the FINA RV O with
O2 M 15x1 - G 3/8“ connection tube.
Connection to FINA AIR precision regulation valve
The flow of volume supplied to the hand-held nebuliser is regulated using the FINA RV A with O2
M 15x1 - M 18x1 connection tube.
Connection to O2 pressure reducing valve with flowmeter:
The 9/16“ - G 3/8“ adapter and the O2 M 15x1 - G 3/8“ connection tube are used for connection
to an O2 pressure reducing valve with flowmeter. The flow of volume supplied to the hand-held
nebuliser is regulated by O2 pressure reducing valve with flowmeter, 0-15 l/min:
Accessories for inhalation:
• 9/16“ - G 3/8“ adapter (REF 5752 2745)
• 9/16“ - M 18x1 adapter (REF 5752 2762)
• O2 ISO M 15x1 - G 3/8“ connection tube (REF 5752 5067)
• 15x1 -M 18x1 AIR FN connection tube (REF 5752 5032)
• M 15x1 -M 18x1 AIR ISO connection tube (REF 5752 5033)
• Medication cup (REF 5750 6210)