Initial operation
Equipment inspection
GA 5750.4172 EN 14
15
3
3
Initial operation
3.1
Equipment inspection
DANGER!
Equipment inspection!
Only components which are in perfect condition can ensure proper functioning of
the product.
The components will thus have to be carefully inspected before using the unit.
WARNING!
Infection hazard!
Contaminated components may endanger the health of the staff and the patients.
Ensure the product is prepared as per hygiene standards before using it for the first
time.
Ensure that the unit has been properly cleaned and that there are no residues or soiling.
Ensure proper functionality of the unit.
Do not use damaged parts.
Ensure that the thread (M 15x1) of the handheld nebuliser matches the thread of the
connection tube to be connected.
Ensure that the thread (G 3/8“ / M 18x1) of the fine regulating valve, flowmeter or pressure
reducing valve with flowmeter matches the connection tube to be connected.
3.2
Assembly of hand-held nebuliser
7
8
6
4
9
1
2
3
5
Fig. 2: Assembly of hand-held nebuliser
Insert the groove of the nozzle (1) into the
gap of the suction sleeve (2).
The nozzle engages with an audible
click.
Bore holes of suction sleeve and nozzle
are congruent one on top of the other.
Plug the piece of tube (3) onto the suction
sleeve.
Insert profile gasket (4) into nebulizer
housing (5).
Screw medicine cup (6) onto nebulizer
housing.
Plug adapter (7) onto inhalation mask (8).
Plug inhalation mask with adapter onto
connector (9).
3.3
Filling the hand-held nebuliser
Unscrew the medicine cup from the nebuliser housing.
Hold medicine cup with one hand and fill with medication.
Screw the medicine cup onto nebuliser housing and hand to patient or suspend with fastening
chain.