3.2 Operating Environment
The following conditions are necessary to ensure correct instrument operation and accurate results:
•
Place the instrument firmly on an even surface. Do not place any objects above or under the instrument.
•
Maintain an ambient temperature of 5ºC to 30ºC for proper instrument operation.
•
The relative humidity should be less than 90%; moisture should not condense on the instrument.
•
Leave at least a 15 cm (6 in.) clearance at the top and on all sides for air circulation to avoid overheating
of electrical parts.
•
Do not operate or store the instrument in extremely dusty, damp or wet locations.
•
Keep the surface of the instrument, the cell compartment, and all accessories clean and dry. Splashes or
spills on and in the instrument should be cleaned up immediately.
Important Note:
Protect the instrument from temperature extremes, including heaters, direct sunlight, and other
heat sources.
3.3 Power Connections
Plug the power supply into the appropriate connector on the back panel of the analyzer, then plug the power
cord into a power outlet (100 – 240 V~ / 45 – 60 Hz). Touch the red power key on the top right of the analyzer to
turn the instrument on.
3.4 Battery Power
The L100 can also be powered using a battery pack (intended for portable field use). The battery pack is plugged
into the back panel of the analyzer.
Please refer to section 5.2 5.2 Battery
Pack
for more information on battery
charging
.
3.5 Personal Computer (PC) Connection
The L100 has one USB interface located on the back of the analyzer. This is used for communication with a PC.
The software is supplied preinstalled on laptop PC and is available for purchase from MANTECH or one of its
authorized distributors.
Note
:
USB cables must not be longer than 3 meters (10 feet).
4. System Operations
4.1 Overview
The following are the steps involved in setting up the L100 analyzer for sample analysis:
1.
Preparation of solutions (calibration/reference, blank and sample).
2.
Priming lines and analyzer assembly.
3.
Pre-conditioning of sensor/calibration.
4.
PeCOD® menu structure