Ensure you prepare sufficient amounts of each solution, taking into account the volume required for priming
(~2mL per prime), calibration and sample analysis. It is recommended to prepare a total minimum volume of
20mL for each sample, and 250mL and 500mL of pre-mixed calibrant and blank, respectively.
4.4 Priming of Lines
Always prime both Port A and Port B lines 3 times before any analysis to ensure that both lines are filled with
liquid and there are no air bubbles present. This can be confirmed by visually checking all flow lines. For
automated system, 6 primes of Port A are required to accommodate the longer sampling line.
To prime the lines manually from the touchpad:
1.
Press
MENU
.
2.
Arrow down to select
Operation
and press
ENTER
.
3.
In the Operation menu, select
Prime Lines
and press
ENTER
.
4.
In the Prime Lines menu, select
Prime Port A
and press
ENTER
. The solution will flow in through the
tube connected into Port A and flow out through the tube connected to Port W. The analyzer will beep
twice at the end. Repeat two more times.
5.
In the Prime Lines menu, arrow down to select
Prime Port B
and press
ENTER
.
The solution will flow in through the tube connected into Port B and flow out through the tube
connected to Port W. The analyser will beep twice at the end. Repeat two more times.
4.5 New Sensor Routine and Calibration
Important Note:
It is recommended that the system be calibrated a minimum of two times each day it is in use
to ensure that the sensor is fully conditioned before running samples. The “New Sensor Routine” can be used to
speed up the process of “breaking-in” a new sensor.
To run a new sensor routine manually from the touchpad:
1.
Press
MENU
.
2.
Arrow down to select
Set up
and press
ENTER
.
3.
In the Set up menu, arrow down to select
Sensor Operation
and press
ENTER
.
4.
Select
New Sensor
and press
ENTER
.
5.
At the prompt “Confirm new sensor?”, press
ENTER
. The PeCOD® will now begin the new sensor
routine, which comprises of a normalization and burn-in phase of Port B.
6.
After 2.5 minutes, the analyzer will beep and ask the user to verify the COD range. Press
ENTER
to
confirm the range. The PeCOD® will now do a full calibration.
7.
Partway through the calibration, the system will beep again. Press
ENTER
to continue.