21
Error
Group
Sub
Code
Name
Description of Problem
Suggested actions
problem, please contact
MANTECH or your local
MANTECH authorised
representative.
8
4
Sol’n not
Presented
User did not press enter upon
system prompt.
•
Restart calibration and ensure
to press enter within 6 minutes
after prompt to avoid timeout.
11
2
Analyser Lid
Open
Lid is open or not securely
latched.
•
Make sure analyser lid is closed
and correctly secured via the
front latch.
11
4
Vaux out of
range
Auxiliary voltage is over range
•
Make sure there are no
bubbles present in the line;
prime lines, and then re-run
analysis.
•
The current and voltage
applied may be too high, try
decreasing the set baseline
and recalibrate.
•
Check electrical contacts
between the connection pins
on analyser board with the
electrode block. Very gently
clean if necessary using
isopropanol and a lint free
cloth.
•
If the above do not rectify the
problem, the sensor or
electrode block may need to
be replaced.
11
6
I
work
out of
range
Current is overrange
•
Make sure there are no
bubbles present in the line;
prime lines, and then re-run
analysis.
•
The current and voltage
applied may be too high, try
decreasing the set baseline
and recalibrate.
•
Check electrical contacts
between the connection pins
on analyser board with the
electrode block. Very gently