The PeCOD
®
L100 analyzer gives a direct measurement of the oxidation of organic compounds, thus providing a
real measurement of organic pollutants and not an inferred one. The reaction produces mainly carbon dioxide
& water from organic C, H and O.
1.3 Technical and General Specifications
* Specifications are subject to change without notice
ANALYSIS DATA
Oxidation process
Photocatalytic oxidation
Catalyst
Titanium Dioxide (TiO
2
)
Method Detection Limit
*0.7ppm
Reproducibility
+/- 10%
Light source
UV LED
*MDL determined for Advanced Blue range
GENERAL SPECIFICATIONS
Construction
ABS
Dimensions (approximate L x W x H)
370mm x 230mm x 240mm
Weight
< 5 kg
Security
4 digit pin (optional)
Parameter(s)
COD (unit of measurement: ppm or mg/L)
BOD (ppm or mg/L)
Measurement
Dilution ≤ 15,000 ppm
ELECTRICAL AND STANDARDS CONFORMITY
Power Requirements
100V to 240V AC / 45 to 65 Hz INPUT – 24V DC X
Amp OUTPUT, Optional battery pack for field use
Current Consumption
2.0 A (maximum)
Enclosure
IP55 to EN60529
Protection Class
Continuous short circuit protection. CEC Level IV
compliant. UL compliant
EMC Emission and Noise Immunity
EN61326-1:2006
FCC Part 15 Subparts A and B
Certification
CE. FCC