MAN-MD-H-0507-01
Page
15
of
24
5.3.2 Auto Reading
The Auto reading function can be activated from 1 to 250 seconds,this will set the time between
readings.
NOTE: If you configure the Auto reading for 5 seconds the unit will make readings every 5
seconds until it is turned off.
Using ▲ or ▼ change the desired time between readings, press and hold
SAVE
for 3 seconds
to store the data and
ESC
to return to the previous menu.
5.3.3 Color Compensation
The instrument can compensate for the color the sample for a more acurate reading.
Using ▲ or ▼ select yes or no, when done, press and hold
SAVE
for 3 seconds to store the
data and
ESC
to return to the previous menu.
NOTE: When activated (Yes selected) “C” will apear in the upper right corner of the display in
the reading mode screen.
5.3.4 Hab. Tests
You can define which courves will appear in the courve selection menu (when you press and
release the Menu key).
Press
ESC
to remove the * icon from the courves you don´t want and
READ
to put the * icon in
the ones you want.
Press and hold
SAVE
for 3 seconds to store the data and
ESC
to return to the previous menu.
5.3.5
Measuring Mode
Here you can select the type of measurent you want to take frmo, fast setling, Media or Normal.
Using ▲ or ▼ select yes or no, when done, press and hold
SAVE
for 3 seconds to store the
data and
ESC
to return to the previous menu.
Fast Setling -When selected, instrument will take a snpshot of the sample and display the
immediate reading before particles settle in the vial. (for high solids samples)
Normal - mediana
Media
– calculates de media of the readings