background image

MAN-MD-H-0507-01 

 

Page 

14

 of 

24

 

 

 

5.2.2 Backlight Time 

 

From 0 up to 60 minutes of backlight on. 
 

Using ▲ or ▼ change the time to the desired backlight time, when done, press and hold 

SAVE 

for 3 seconds to store the data and 

ESC 

to return to the previous menu. 

 

5.2.3 Backlight Level 

 

From 0 up to 100 (intensity level) 

 

Using ▲ or ▼ change the level to the desired, when done, press and hold 

SAVE 

for 3 seconds 

to store the data and 

ESC 

to return to the previous menu. 

 

5.2.4 Partial Res. 

 

Wen this mode is On, the unit will show resoults during the measurement cycle, intermediate 
readings. 

 

Using ▲ or ▼ change to select yes or no, when done, press and hold 

SAVE 

for 3 seconds to 

store the data and 

ESC 

to return to the previous menu. 

 

5.2.5 Big number 

 

Using ▲ or ▼ change it from yes or no, when done, press and hold 

SAVE 

for 3 seconds to store 

the data and 

ESC 

to return to the previous menu. 

 

If you shoose yes, the numbers will apear larger on the display 

 

5.3 Instrument 

 

When inside this configuration you can set Auto off, Readings, Color compensation, courves, fast 
settling, Sampling, ID, Calibratio int

erval, personalization, patrimony, use ▲ or ▼ to select between 

the options and 

READ 

to enter it or 

ESC 

to go gack to the previous menu. 

 

5.3.1 Auto off 

 

The Auto off function shall be activated to save the bateryes, iit can be configurated to turn the 
unit off after 0 to 60 minutes of inactivity. 

 

Using ▲ or ▼ change the time to the desired level, when done, press and hold 

SAVE 

for 3 

seconds to store the data and 

ESC 

to return to the previous menu. 

 

NOTE: When the time is in 0 minutes the auto off will not be operational. 

Содержание Ap2000

Страница 1: ...MAN MD H 0507 01 Page 1 of 24 Instruction Manual MANTECH Ap2000 Turbidimeter...

Страница 2: ...tation direct incidental and consequential damages and disclaims such damages to the full extent permitted under applicable law The user is solely responsible to identify critical application risks an...

Страница 3: ...data Auto ShutOff Programmable from 1 to 60 min Fast Cal function Quick Calibration for single point Software Functions Signal averaging Fast Settling results freezing analyst and sample identificatio...

Страница 4: ...n Philips screw driver remove the battery cover 2 Install 4 AA alkaline or 4 AA nickel metal hydride NiMH batteries Make sure that the batteries are installed in the correct orientation 3 Replace the...

Страница 5: ...to be USB transfered and selects the parameters 4 ESC Powers off the unit aborts operations return to the previous screen 5 ON Off Read switches the instrument on confirms options sample reading move...

Страница 6: ...i e Access the Service Functions only for certified technicians Please see Menu structure in the last page of this manual Press and hold the MENU key for 3 seconds to enter the Menu Function You will...

Страница 7: ...User1 User calibrated courve require full calibration of this courve User2 User calibrated courve require full calibration of this courve Press and release save to use the desired measuring range fun...

Страница 8: ...the list press and hold SAVE for 3 seconds to store the data 4 3 1 USer Select the User for the sample Use and to choose an user from the list press and hold SAVE for 3 seconds to store the data 4 3 2...

Страница 9: ...VIAL IN THE READING CELL Align the l mark of the sample vial forward with the arrow on the sample cup of the instrument and close the instrument lid If using Auto mode just wait for the reading to sta...

Страница 10: ...l must be aligned with the arrow in the instrument sample chamber Sample Vials calibration The turbidimeter AP 2000 vials are factory selected for optimal precision however in low turbidity conditions...

Страница 11: ...per 4 5 1 Guided Cal 1 Press and Hold MENU key for 3 seconds You will see ID Identification Use the or keys to select Calibrate then press READ to enter it 2 The display will show guided cal press REA...

Страница 12: ...s to the Calibrate menu 5 After calibrating the 0 1 NTU standard perform the calibration of the other standards 20 NTU 100 NTU and 800 NTU Recalibrate the 0 1 NTU for better accuracy If you have chose...

Страница 13: ...e menu structure in the last page of this manual Using or you can navigate between all the configurable functions When you reach the desired function press READ to enter it or or ESC to go ack to the...

Страница 14: ...the previous menu 5 2 5 Big number Using or change it from yes or no when done press and hold SAVE for 3 seconds to store the data and ESC to return to the previous menu If you shoose yes the numbers...

Страница 15: ...lected C will apear in the upper right corner of the display in the reading mode screen 5 3 4 Hab Tests You can define which courves will appear in the courve selection menu when you press and release...

Страница 16: ...IN on the Display use to scroll faster to letters to scroll faster to numbers both can be used to go up or down and MENU READ send the cursor to the right left press and hold SAVE for 3 seconds to sto...

Страница 17: ...3 seconds to store the data Press and Hold ESC for 3 seconds to return to the previous menu 5 3 7 2 2 Request To define when the sample name will be required Choose between the following options using...

Страница 18: ...ct tha name you want to give to the unit Using to scroll to letters to scroll to numbers both can be used to go up or down and MENU READ send the cursor to the right left press and hold SAVE for 3 sec...

Страница 19: ...ou can select to send all measurements only part of them Using or select Auto Manual and Off when done press and hold SAVE for 3 seconds to store the data and ESC to return to the previous menu Auto S...

Страница 20: ...t in order to have full confidence in instrument performance For more information on how to calibrate user courve please refer to the advanced operations instructions manual for the turbidity meter AP...

Страница 21: ...tween 0 and 5 press and hold SAVE for 3 seconds to store the data 5 6 2 2 Password Using to scroll faster to letters to scroll faster to numbers both can be used to go up or down and MENU READ send th...

Страница 22: ...d 5 press and hold SAVE for 3 seconds to store the data 5 6 4 2 Password Using to scroll faster to letters to scroll faster to numbers both can be used to go up or down and MENU READ send the cursor t...

Страница 23: ...Datalog Visualize Log Transmit Diagnostic Signal Sensor Batery Duty Current mA Light Temperature Blank NL_Blank F Scale NL_F Scale Readings Reset Calibr Defaut Activate Dafault Save Defaut Light Cal...

Страница 24: ...0507 01 Page 24 of 24 Document Change Log Version Date Author Changes 1 7 Aug 2019 Laura Martin Document Created www mantech inc com MANTECH Inc 5473 Highway 6 North Guelph Ontario N1H 6J2 Canada 519...

Отзывы: