42
PMA1000/PMA1000L | 1.00.00
www.mc-techgroup.com
By tapping on the items of the list, you will reach the corresponding screen to enter the
settings.
1
2
Fig. 42: Channel-specific settings list
1
Selection button
2
Channel-specific settings list
Here is a list of the parameters which belong to the ‘1 = Channel K1-Kn settings’.
Channel Parameter description
Default
value
Selection: 1= Basic settings
K1
channel ID
PMA
K1
display number 1 - 10
1
K1
op. Lim1 [vol%]
- (20.000*)
K1
op. Lim2 [vol%]
- (18.000*)
K1
op. mode Lim1 0: inactive, 1: <, 2: ≤, 3: >, 4: ≥
0 (1*)
K1
op. mode Lim2 0: inactive, 1: <, 2: ≤, 3: >, 4: ≥
0 (1*)
K1
concentration averaging value: no=0, yes=1..100
0
K1
unit temperature (1 = °C, 2 = °F]
1
K1
unit pressure (1 = bar, 2 = hPa, 3 = mbar, 4 = psi)
3
K1
pressure calibration offset P-IN
0.000
K1
pressure calibration offset P-OUT
0.000
K1
threshold pressure [mbar] min
600
K1
threshold pressure [mbar] max
1600
K1
unit sample flow (1 = l/h, 2 = l/min)
1
K1
correction factor sample flow
1.000
K1
threshold digit min
0
K1
threshold digit max
0
K1
threshold digit default
1300
K1
number of decimal digits
2